Literature DB >> 16144845

Dynamic changes in right ventricular pressures during haemodialysis recorded with an implantable haemodynamic monitor.

Frieder Braunschweig1, Barbro Kjellström, Mats Söderhäll, Naomi Clyne, Cecilia Linde.   

Abstract

BACKGROUND: Intermittent and chronic volume overload contributes to the development of cardiovascular disease in patients on maintenance haemodialysis (HD). Continuous monitoring of central haemodynamic parameters may provide valuable information to improve volume control, particularly in patients with left ventricular dysfunction.
METHODS: Five patients on HD, age 53-76 years, with systolic and/or diastolic dysfunction (EF 20-50%) received an implantable haemodynamic monitor (IHM) (Chronicle model 9520, Medtronic). The IHM consists of a memory device implanted subcutaneously and a transveneous right ventricular (RV) lead carrying a pressure sensor. It continuously records heart rate, RV systolic (RVSP) and diastolic pressures (RVDP), RV dP/dt and an estimate of pulmonary artery diastolic pressure (ePAD). Continuous haemodynamic profiles were recorded in all patients.
RESULTS: During dialysis RVSP and ePAD dropped by a mean of 39 and 50%, respectively. RVDP decreased by 6.6 mmHg. The lowest pressures occurred during the first 90 min of dialysis and were partly restored at the end of the procedure. Long-term haemodynamic monitoring unmasked severe volume overload in one patient, when dry weight was kept stable despite a decrease in lean body mass. In another patient with recurrent dyspnea after dialysis, paroxysmal atrial fibrillation, regularly occurring during dialysis, was identified as the cause of symptoms.
CONCLUSION: The implanted haemodynamic monitor was a sensitive indicator for changes in volume load. Continuous haemodynamic monitoring may offer a valuable tool to improve volume management in dialysis patients with left ventricular dysfunction.

Entities:  

Mesh:

Year:  2005        PMID: 16144845     DOI: 10.1093/ndt/gfi145

Source DB:  PubMed          Journal:  Nephrol Dial Transplant        ISSN: 0931-0509            Impact factor:   5.992


  8 in total

Review 1.  The Ebb and Flow of Echocardiographic Cardiac Function Parameters in Relationship to Hemodialysis Treatment in Patients with ESRD.

Authors:  Charalampos Loutradis; Pantelis A Sarafidis; Christodoulos E Papadopoulos; Aikaterini Papagianni; Carmine Zoccali
Journal:  J Am Soc Nephrol       Date:  2018-03-28       Impact factor: 10.121

2.  Estimated pulmonary artery systolic pressure and self-reported physical function in patients on hemodialysis.

Authors:  Adrian P Abreo; Charles A Herzog; Nancy G Kutner; Janice Lea; Kirsten L Johansen
Journal:  Am J Nephrol       Date:  2015-06-13       Impact factor: 3.754

3.  The occurrence of atrial fibrillation in dialysis patients and its association with left atrium volume before and after dialysis.

Authors:  Beata Franczyk; Anna Gluba-Brzózka; Piotr Bartnicki; Jacek Rysz
Journal:  Int Urol Nephrol       Date:  2017-02-25       Impact factor: 2.370

4.  Increased Right Ventricular Pressure as a Predictor of Acute Decompensated Heart Failure in End-Stage Renal Disease Patients on Maintenance Hemodialysis.

Authors:  Bong-Joon Kim; Soo-Jin Kim; Sung-Il Im; Hyun-Su Kim; Jung-Ho Heo; Ho Sik Shin; Ye Na Kim; Yeonsoon Jung; Hark Rim
Journal:  Int J Heart Fail       Date:  2022-04-15

5.  Cardiovascular outcome trials in patients with chronic kidney disease: challenges associated with selection of patients and endpoints.

Authors:  Patrick Rossignol; Rajiv Agarwal; Bernard Canaud; Alan Charney; Gilles Chatellier; Jonathan C Craig; William C Cushman; Ronald T Gansevoort; Bengt Fellström; Dahlia Garza; Nicolas Guzman; Frank A Holtkamp; Gerard M London; Ziad A Massy; Alexandre Mebazaa; Peter G M Mol; Marc A Pfeffer; Yves Rosenberg; Luis M Ruilope; Jonathan Seltzer; Amil M Shah; Salim Shah; Bhupinder Singh; Bergur V Stefánsson; Norman Stockbridge; Wendy Gattis Stough; Kristian Thygesen; Michael Walsh; Christoph Wanner; David G Warnock; Christopher S Wilcox; Janet Wittes; Bertram Pitt; Aliza Thompson; Faiez Zannad
Journal:  Eur Heart J       Date:  2019-03-14       Impact factor: 29.983

Review 6.  Therapeutic and diagnostic role of electrical devices in acute heart failure.

Authors:  Frieder Braunschweig
Journal:  Heart Fail Rev       Date:  2007-06       Impact factor: 4.654

Review 7.  Evolution from electrophysiologic to hemodynamic monitoring: the story of left atrial and pulmonary artery pressure monitors.

Authors:  Deirdre M Mooney; Erik Fung; Rahul N Doshi; David M Shavelle
Journal:  Front Physiol       Date:  2015-10-07       Impact factor: 4.566

8.  Dialysis-specific factors and incident atrial fibrillation in hemodialysis patients.

Authors:  Seung Don Baek; Soomin Jeung; Jae-Young Kang; Ki Hyun Jeon
Journal:  Ren Fail       Date:  2020-11       Impact factor: 2.606

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.