Variability among five over-the-counter blood glucose monitors.

Mary M Kimberly1, Hubert W Vesper, Samuel P Caudill, Steven F Ethridge, Enada Archibold, Kimberly H Porter, Gary L Myers.   

Abstract

BACKGROUND: The American Diabetes Association recommends that people with diabetes use self-monitoring to control their blood glucose concentration. To assess the need for standardization, we evaluated the variability among 5 of the most common monitors: MediSense Precision Xtra, Ascencia Dex, Prestige Smart System, OneTouch Ultra, and Accu-Chek Advantage.
METHODS: We took steps to minimize preanalytical variation. We also eliminated user variability by using one trained operator to collect samples and perform all testing. Each monitor was used twice with each participant; one test was performed using an aged strip and the other using a fresh strip. We compared monitors using a separate ANOVA for each concentration range and strip lot.
RESULTS: The total CVs and the within-strip lot CVs were not statistically different among monitors, ranging from 3.1% to 11.3% and from 2.1% to 8.5%, respectively. There were statistically significant differences among monitors for among-strip lot CVs, which ranged from nearly 0% to 7.5%. The degree of significance increased as the concentration range increased [3.9-5.5 mmol/l: p<0.05; 5.6-7.7 mmol/l: p =0.003; 7.8-11.1 mmol/l: p < 0.001]. The average percent difference between monitor pairs was statistically significant (p < 0.05) in more than half of the paired comparisons, with significant differences ranging from 5.7% to 32.0%.
CONCLUSIONS: Monitor results can vary significantly so that agreement among them is poor. Standardization is necessary to minimize variability and to improve patient care.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16143321     DOI: 10.1016/j.cca.2005.07.027

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  14 in total

1.  New Criteria for Assessing the Accuracy of Blood Glucose Monitors meeting, October 28, 2011.

Authors:  John Walsh; Ruth Roberts; Robert A Vigersky; Frank Schwartz
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

2.  System Accuracy Evaluation of Four Systems for Self-Monitoring of Blood Glucose Following ISO 15197 Using a Glucose Oxidase and a Hexokinase-Based Comparison Method.

Authors:  Manuela Link; Christina Schmid; Stefan Pleus; Annette Baumstark; Delia Rittmeyer; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2015-04-14

3.  Approaches for improving glucose monitor measurements for self-monitoring of blood glucose: from measurement harmonization to external quality assessment programs.

Authors:  Hubert W Vesper; Gary L Myers
Journal:  J Diabetes Sci Technol       Date:  2007-03

4.  Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197.

Authors:  Annette Baumstark; Stefan Pleus; Christina Schmid; Manuela Link; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

5.  Evaluation of the analytical performance of the coulometry-based Optium Omega blood glucose meter.

Authors:  Bogdan Solnica; Beata Kusnierz-Cabala; Krystyna Slowinska-Solnica; Przemyslaw Witek; Agnieszka Cempa; Maciej T Malecki
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

6.  Blood sugar measurement in zebrafish reveals dynamics of glucose homeostasis.

Authors:  Stefani C Eames; Louis H Philipson; Victoria E Prince; Mary D Kinkel
Journal:  Zebrafish       Date:  2010-06       Impact factor: 1.985

Review 7.  Analytical and clinical performance of blood glucose monitors.

Authors:  Suzanne Austin Boren; William L Clarke
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

8.  Finger-stick glucose monitoring: issues of accuracy and specificity.

Authors:  Leann Olansky; Laurence Kennedy
Journal:  Diabetes Care       Date:  2010-04       Impact factor: 19.112

9.  Glucose measurement: time for a gold standard.

Authors:  Joakim Hagvik
Journal:  J Diabetes Sci Technol       Date:  2007-03

10.  Validity and Reliability of a Glucometer Against Industry Reference Standards.

Authors:  Amanda J Salacinski; Micah Alford; Kathryn Drevets; Sarah Hart; Brian E Hunt
Journal:  J Diabetes Sci Technol       Date:  2014-01-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.