Literature DB >> 16139707

Polyethylene liner failure in second-generation Harris-Galante acetabular components.

Byung-Woo Min1, Kwang-Soon Song, Chul-Hyung Kang, Ye-Yeon Won, Kyung-Hoi Koo.   

Abstract

Polyethylene liner failure of cementless acetabular components has been reported to be the most common complication of current modular cups. We performed a retrospective review of 128 consecutive primary total hip arthroplasties in 111 patients who had a second-generation Harris-Galante cup (Harris-Galante II, Zimmer, Warsaw, Ind). The mean length of the follow-up period was 8.0 years (range 5-12.9 years), with 26 cups (20.2%) having undergone revision by that time. Twenty-two hips (17.3%) had failure of the polyethylene liner. We found that polyethylene liner failure remains the main cause of revision of this modular cup. The main difference between our population of patients and other populations that have used this cup and liner is our population's cultural inclination toward squatting and leg crossing, body positions that increased the risk of impingement and eccentric wear.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16139707     DOI: 10.1016/j.arth.2004.11.006

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  5 in total

1.  Distributed analysis of hip implants using six national and regional registries: comparing metal-on-metal with metal-on-highly cross-linked polyethylene bearings in cementless total hip arthroplasty in young patients.

Authors:  Ove Furnes; Elizabeth Paxton; Guy Cafri; Stephen Graves; Barbara Bordini; Thomas Comfort; Moises Coll Rivas; Samprit Banerjee; Art Sedrakyan
Journal:  J Bone Joint Surg Am       Date:  2014-12-17       Impact factor: 5.284

2.  Second-generation modular acetabular components provide fixation at 10 to 16 years.

Authors:  Paul F Lachiewicz; Elizabeth S Soileau
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

3.  Wear versus thickness and other features of 5-Mrad crosslinked UHMWPE acetabular liners.

Authors:  Fu-Wen Shen; Zhen Lu; Harry A McKellop
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

Review 4.  Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survivalstudies.

Authors:  Bart G Pijls; Marc J Nieuwenhuijse; Marta Fiocco; Josepha Wm Plevier; Saskia Middeldorp; Rob Ghh Nelissen; Edward R Valstar
Journal:  Acta Orthop       Date:  2012-11-05       Impact factor: 3.717

5.  Catastrophic Failure of the Acetabular Polyethylene Liner in Ceramic-on-Polyethylene Total Hip Arthroplasty.

Authors:  Givenchy Manzano; Rayna A C Levin; Michael B Mayor; Ran Schwarzkopf
Journal:  J Orthop Case Rep       Date:  2014 Apr-Jun
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.