Comparison of hydrophobic heat and moisture exchangers with heated humidifier during prolonged mechanical ventilation.

J P Roustan1, J Kienlen, P Aubas, S Aubas, J du Cailar.   

Abstract

Inspired gases must be warmed and humidified during mechanical ventilation. In a prospective randomized study we compared the performance of a heated humidifier (HH) (Draegger Aquaport) and a heat and moisture exchanger (HME) (Pall Filter BB 2215). A total of 116 patients requiring mechanical ventilation (Servo 900 C Siemens) were enrolled into the study and were randomly assigned to 2 groups. Patients in group I were ventilated with a traditional breathing circuit with HH and patients in group II using a simplified circuit with HME. Pre-existing and hospital acquired atelectasis and pneumonia, occurrence of endotracheal tube (ET) occlusion and ventilatory parameters (respiratory rate, tidal volume) were studied. No statistical difference was found between groups for each parameter except the greater frequency of ET occlusions in the II group (0/61 vs 9/55) (p = 0.0008). Pall Filter (PF), a hydrophobic filter, humidifies the dry gases from the condensed water which is put down on the HME surfaces during cooling of saturated expired gases. This purely physical property is linked to the magnitude of the thermic gradient between the expired gases and the ambiant temperature. Performance impairment of PF in our study might be due to high ambiant temperature in the intensive care unit (usually around 28 degrees C) which reduces thermic gradient and water exchanges. We conclude that efficiency of PF may be weak in some conditions of ambiant temperature.

Entities:  

Mesh:

Year:  1992        PMID: 1613206     DOI: 10.1007/bf01705040

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  24 in total

1.  Assessment of condenser-humidifiers with special reference to a multiplegauze model.

Authors:  W W MAPLESON; J G MORGAN; E K HILLARD
Journal:  Br Med J       Date:  1963-02-02

2.  Endotracheal tube occlusion associated with the use of heat and moisture exchangers in the intensive care unit.

Authors:  I L Cohen; P F Weinberg; I A Fein; G S Rowinski
Journal:  Crit Care Med       Date:  1988-03       Impact factor: 7.598

3.  The use of bolus normal saline instillations in artificial airways: is it useful or necessary?

Authors:  M H Ackerman
Journal:  Heart Lung       Date:  1985-09       Impact factor: 2.210

4.  Complications of assisted ventilation. A prospective study of 354 consecutive episodes.

Authors:  C W Zwillich; D J Pierson; C E Creagh; F D Sutton; E Schatz; T L Petty
Journal:  Am J Med       Date:  1974-08       Impact factor: 4.965

5.  An assessment of methods of humidification of inspired gas.

Authors:  B Hayes; J S Robinson
Journal:  Br J Anaesth       Date:  1970-02       Impact factor: 9.166

6.  Temperature variations in disposable heat and moisture exchangers.

Authors:  B Revenäs; C E Lindholm
Journal:  Acta Anaesthesiol Scand       Date:  1980-06       Impact factor: 2.105

7.  Contaminated condensate in mechanical ventilator circuits. A risk factor for nosocomial pneumonia?

Authors:  D E Craven; T A Goularte; B J Make
Journal:  Am Rev Respir Dis       Date:  1984-04

8.  Moisture-conserving efficiency of condenser humidifiers.

Authors:  M Ogino; R Kopotic; F L Mannino
Journal:  Anaesthesia       Date:  1985-10       Impact factor: 6.955

9.  Heat and moisture exchangers and vaporizing humidifiers in the intensive care unit.

Authors:  C Martin; G Perrin; M J Gevaudan; P Saux; F Gouin
Journal:  Chest       Date:  1990-01       Impact factor: 9.410

10.  Contamination control in long-term ventilation. A clinical study using a heat- and moisture-exchanging filter.

Authors:  J Gallagher; J E Strangeways; J Allt-Graham
Journal:  Anaesthesia       Date:  1987-05       Impact factor: 6.955

View more
  19 in total

Review 1.  Efficacy of heat and moisture exchangers in preventing ventilator-associated pneumonia: meta-analysis of randomized controlled trials.

Authors:  Axel Kola; Tim Eckmanns; Petra Gastmeier
Journal:  Intensive Care Med       Date:  2004-09-11       Impact factor: 17.440

2.  Ventilator-associated pneumonia: diagnosis, treatment, and prevention.

Authors:  Steven M Koenig; Jonathon D Truwit
Journal:  Clin Microbiol Rev       Date:  2006-10       Impact factor: 26.132

3.  Under-humidification and over-humidification during moderate induced hypothermia with usual devices.

Authors:  François Lellouche; Siham Qader; Solenne Taille; Aissam Lyazidi; Laurent Brochard
Journal:  Intensive Care Med       Date:  2006-05-23       Impact factor: 17.440

4.  Comparison of the effects of two humidifier systems on endotracheal tube resistance.

Authors:  Indalecio Morán; Belén Cabello; Elizabeth Manero; Jordi Mancebo
Journal:  Intensive Care Med       Date:  2011-09-27       Impact factor: 17.440

5.  Performance of a hydrophobic heat and moisture exchanger at different ambient temperatures.

Authors:  M Croci; A Elena; M Solca
Journal:  Intensive Care Med       Date:  1993       Impact factor: 17.440

6.  An experimental set-up to test heat-moisture exchangers.

Authors:  N Unal; J C Pompe; W P Holland; I Gültuna; P E Huygen; K Jabaaij; C Ince; B Saygin; H A Bruining
Journal:  Intensive Care Med       Date:  1995-02       Impact factor: 17.440

Review 7.  Humidification and heating of inhaled gas in patients with artificial airway. A narrative review.

Authors:  Gustavo Adrián Plotnikow; Matias Accoce; Emiliano Navarro; Norberto Tiribelli
Journal:  Rev Bras Ter Intensiva       Date:  2018-03

Review 8.  Paradoxical ventilator associated pneumonia incidences among selective digestive decontamination studies versus other studies of mechanically ventilated patients: benchmarking the evidence base.

Authors:  James C Hurley
Journal:  Crit Care       Date:  2011-01-07       Impact factor: 9.097

9.  Evaluating humidity recovery efficiency of currently available heat and moisture exchangers: a respiratory system model study.

Authors:  Jeanette Janaina Jaber Lucato; Alexander Bernard Adams; Rogério Souza; Jamili Anbar Torquato; Carlos Roberto Ribeiro Carvalho; John J Marini
Journal:  Clinics (Sao Paulo)       Date:  2009       Impact factor: 2.365

10.  Ventilatory changes during the use of heat and moisture exchangers in patients submitted to mechanical ventilation with support pressure and adjustments in ventilation parameters to compensate for these possible changes: a self-controlled intervention study in humans.

Authors:  Jeanette Janaina Jaber Lucato; Thiago Marraccini Nogueira da Cunha; Aline Mela Dos Reis; Patricia Salerno de Almeida Picanço; Renata Cléia Claudino Barbosa; Joyce Liberali; Renato Fraga Righetti
Journal:  Rev Bras Ter Intensiva       Date:  2017 Apr-Jun
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.