| Literature DB >> 16130656 |
Conny Danielsson1, Karin Wiberg, Peter Korytár, Sture Bergek, Udo A Th Brinkman, Peter Haglund.
Abstract
Trace analysis of 2,3,7,8-polychlorinated dibenzo-p-dioxins (PCDDs), dibenzofurans (PCDFs) and the 12 WHO-PCBs (four non-ortho and eight mono-ortho congeners that have been assigned toxic equivalence factors, TEFs, by the World Health Organisation) was conducted by comprehensive two-dimensional gas chromatography with a micro electron-capture detector (GC x GC-microECD). Four food matrices (fish oil from herring, spiked cows' milk, vegetable oil and an eel extract) were analysed by two GC x GC laboratories, and four GC-HRMS laboratories generated reference values. The two GC x GC laboratories used different column combinations for separating the target analytes. For the first dimension, non-polar DB-XLB and VF-1 columns were used, and for the second dimension, an LC-50 liquid crystalline column with unique selectivity for planar compounds. The congener-specific and total toxic equivalence (TEQ) data obtained using DB-XLB x LC-50 were in good agreement with results obtained by the GC-HRMS laboratories. The WHO-PCB data obtained with the VF-1 x LC-50 combination was also good, but the PCDD/F concentrations were sometimes overestimated due to matrix interferences. GC x GC-microECD using DB-XLB x LC-50 seems to fulfil the European Community requirements of a screening method for PCDD/F and WHO-PCB TEQ in food.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16130656 DOI: 10.1016/j.chroma.2004.11.071
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759