| Literature DB >> 16126517 |
Abstract
The stable carbon isotopic composition of dissolved inorganic carbon (delta13C(DIC)) is traditionally determined using either direct precipitation or gas evolution methods in conjunction with offline gas preparation and measurement in a dual-inlet isotope ratio mass spectrometer. A gas evolution method based on continuous-flow technology is described here, which is easy to use and robust. Water samples (100-1500 microl depending on the carbonate alkalinity) are injected into He-filled autosampler vials in the field and analysed on an automated continuous-flow gas preparation system interfaced to an isotope ratio mass spectrometer. Sample analysis time including online preparation is 10 min and overall precision is 0.1 per thousand. This method is thus fast and can easily be automated for handling large sample batches.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16126517 DOI: 10.1080/10256010500230023
Source DB: PubMed Journal: Isotopes Environ Health Stud ISSN: 1025-6016 Impact factor: 1.675