Accuracy, reliability, and validity of a spatiotemporal gait analysis system.

Susan Barker1, Rebecca Craik, William Freedman, Nira Herrmann, Howard Hillstrom.   

Abstract

BACKGROUND: Making evidence-based decisions in the clinical management of gait dysfunction requires the ability to measure spatiotemporal gait variables validly and reliably. This study examined the accuracy, reliability, and validity of a clinical gait analysis system that is relatively inexpensive and portable, the GaitMat II (GM).
METHODS: The trustworthiness of measures taken with the GM was investigated by comparing the placement of the switches on three different GM systems. Measures taken with the GM system were compared with the same measures taken simultaneously with the Vicon motion analysis system, using a repetitive dynamic signal presented to both systems.
FINDINGS: Variability was present in the placement of switches on the surfaces of different GM systems. The variability in switch placement was much less than the normal variability of most gait variables measured with the GM. Excellent agreement was found between the GM and Vicon systems for timing variables but poor agreement was found for distance variables. The mean difference between the GM and Vicon measures for distance variables would have little clinical significance in adults for distance measures such as step length or stride length.
CONCLUSION: The results of these studies support the reliability and validity of the GaitMat II as a clinical gait analysis tool.

Entities:  

Mesh:

Year:  2005        PMID: 16122966     DOI: 10.1016/j.medengphy.2005.07.017

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  18 in total

1.  Gait biomechanics, spatial and temporal characteristics, and the energy cost of walking in older adults with impaired mobility.

Authors:  David M Wert; Jennifer Brach; Subashan Perera; Jessie M VanSwearingen
Journal:  Phys Ther       Date:  2010-05-20

2.  Characteristics of walking, activity, fear of falling, and falls in community-dwelling older adults by residence.

Authors:  David M Wert; Jaime B Talkowski; Jennifer Brach; Jessie VanSwearingen
Journal:  J Geriatr Phys Ther       Date:  2010 Jan-Mar       Impact factor: 3.381

3.  Gait variability in older adults: observational rating validated by comparison with a computerized walkway gold standard.

Authors:  Wen-Ni Wennie Huang; Jessie M VanSwearingen; Jennifer S Brach
Journal:  Phys Ther       Date:  2008-08-21

4.  Interpreting the need for initial support to perform tandem stance tests of balance.

Authors:  Elizabeth S Hile; Jennifer S Brach; Subashan Perera; David M Wert; Jessie M VanSwearingen; Stephanie A Studenski
Journal:  Phys Ther       Date:  2012-06-28

5.  A unique foot-worn device for patients with degenerative meniscal tear.

Authors:  Avi Elbaz; Yiftah Beer; Ehud Rath; Guy Morag; Ganit Segal; Eytan M Debbi; Daniel Wasser; Amit Mor; Ronen Debi
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-05-04       Impact factor: 4.342

6.  Patients with knee osteoarthritis demonstrate improved gait pattern and reduced pain following a non-invasive biomechanical therapy: a prospective multi-centre study on Singaporean population.

Authors:  Avi Elbaz; Amit Mor; Ganit Segal; Yoav Aloni; Yee Hong Teo; Yee Sze Teo; Shamal Das-De; Seng Jin Yeo
Journal:  J Orthop Surg Res       Date:  2014-01-02       Impact factor: 2.359

7.  Deviations in gait metrics in patients with chronic ankle instability: a case control study.

Authors:  Roy Gigi; Amir Haim; Elchanan Luger; Ganit Segal; Eyal Melamed; Yiftah Beer; Matityahu Nof; Meir Nyska; Avi Elbaz
Journal:  J Foot Ankle Res       Date:  2015-01-21       Impact factor: 2.303

8.  Analysis of knee flexion characteristics and how they alter with the onset of knee osteoarthritis: a case control study.

Authors:  Ian McCarthy; Diana Hodgins; Amit Mor; Avi Elbaz; Ganit Segal
Journal:  BMC Musculoskelet Disord       Date:  2013-05-21       Impact factor: 2.362

9.  A non-invasive biomechanical device and treatment for patients following total hip arthroplasty: results of a 6-month pilot investigation.

Authors:  Ganit Segal; Yaron Bar-Ziv; Steven Velkes; Vadim Benkovich; Gilad Stanger; Eytan M Debbi; Ronen Debi; Amit Mor; Avi Elbaz
Journal:  J Orthop Surg Res       Date:  2013-05-21       Impact factor: 2.359

10.  Soft tissue artefacts of the human back: comparison of the sagittal curvature of the spine measured using skin markers and an open upright MRI.

Authors:  Roland Zemp; Renate List; Turgut Gülay; Jean Pierre Elsig; Jaroslav Naxera; William R Taylor; Silvio Lorenzetti
Journal:  PLoS One       Date:  2014-04-18       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.