Hemodynamic changes due to stent placement in bifurcating intracranial aneurysms.

Gador Cantón1, David I Levy, Juan C Lasheras.   

Abstract

OBJECT: The aim of this study was to measure changes in intraaneurysm flow dynamics and mechanical stresses resulting from the placement of Neuroform stents in bifurcating intracranial aneurysm models.
METHODS: A digital particle image velocimetry (DPIV) system was used to measure the pulsatile velocity and shear stress fields within the aneurysm and at the aneurysm neck-parent artery interface. The DPIV system provides an instantaneous two-dimensional measurement of the temporal and spatial variations of the velocity vector field of the flow inside the aneurysm pouch and the parent vessel, providing information on both the temporal and spatial variations of the velocity field during the entire cardiac cycle. The corresponding shear stress field was then computed from the velocity field data. A flexible silicone model of bifurcating intracranial aneurysms was used. Two Neuroform stents with a 60- to 65-microm strut thickness and an 11% metal/artery ratio were placed in a Y-configuration, and measurements were obtained after placing the stents.
CONCLUSIONS: Two three-dimensional vortices of different strengths persisted within the aneurysm during the entire cardiac cycle. The peak velocity and strength of these vortices were reduced after placing the two bifurcating stents. The effect of placing the Neuroform stent across the neck of a bifurcating intracranial aneurysm was shown to reduce the magnitude of the velocity of the jet entering the sac by as much as 11%. Nevertheless, the effect of the stents was particularly noticeable at the end of the cardiac cycle, when the residual vorticity and shear stresses inside the sac were decreased by more than 40%.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16121985     DOI: 10.3171/jns.2005.103.1.0146

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  30 in total

1.  Ruptured Intracanalicular Ophthalmic Artery Aneurysm Treated with Low Profile Flow Diverter Device : Case Report.

Authors:  S Sirakov; A Sirakov; H Tsonev; H Hristov
Journal:  Clin Neuroradiol       Date:  2019-05-22       Impact factor: 3.649

2.  Y-stent-assisted coil embolization of wide-neck intracranial aneurysms. A single center experience.

Authors:  E Akgul; E Aksungur; T Balli; B Onan; D M Yilmaz; S Bicakci; T Erman
Journal:  Interv Neuroradiol       Date:  2011-04-18       Impact factor: 1.610

3.  Stent-assisted coiling of intracranial aneurysms using LEO stents: long-term follow-up in 153 patients.

Authors:  Jacques Sedat; Yves Chau; Jean Gaudart; Marina Sachet; Stephanie Beuil; Michel Lonjon
Journal:  Neuroradiology       Date:  2017-12-23       Impact factor: 2.804

4.  Y-stent-assisted coil embolization of anterior circulation aneurysms using two Solitaire AB devices: a single center experience.

Authors:  M Martínez-Galdámez; P Saura; J Saura; A Martínez; J M De Campos; A Pérez
Journal:  Interv Neuroradiol       Date:  2012-06-04       Impact factor: 1.610

5.  Comparison of 2-year angiographic outcomes of stent- and nonstent-assisted coil embolization in unruptured aneurysms with an unfavorable configuration for coiling.

Authors:  G Hwang; H Park; J S Bang; S-C Jin; B C Kim; C W Oh; H S Kang; M H Han; O-K Kwon
Journal:  AJNR Am J Neuroradiol       Date:  2011-08-18       Impact factor: 3.825

6.  Effects of stenting the parent artery on aneurysm filling and gene expression of various potential factors involved in healing of experimental aneurysms.

Authors:  T Darsaut; I Salazkin; C Ogoudikpe; G Gevry; F Bouzeghrane; J Raymond
Journal:  Interv Neuroradiol       Date:  2007-01-19       Impact factor: 1.610

7.  Double stent-assisted coil embolization treatment for bifurcation aneurysms: immediate treatment results and long-term angiographic outcome.

Authors:  K Yavuz; S Geyik; S Cekirge; I Saatci
Journal:  AJNR Am J Neuroradiol       Date:  2013-03-28       Impact factor: 3.825

8.  Pipeline embolization device in aneurysmal subarachnoid hemorrhage.

Authors:  J P Cruz; C O'Kelly; M Kelly; J H Wong; W Alshaya; A Martin; J Spears; T R Marotta
Journal:  AJNR Am J Neuroradiol       Date:  2012-10-11       Impact factor: 3.825

9.  Effect of flow diverter porosity on intraaneurysmal blood flow.

Authors:  Luca Augsburger; Mohamed Farhat; Philippe Reymond; Edouard Fonck; Zsolt Kulcsar; Nikos Stergiopulos; Daniel A Rüfenacht
Journal:  Klin Neuroradiol       Date:  2009-08-23

10.  Safety and efficacy of neuroform for treatment of intracranial aneurysms: a prospective, consecutive, French multicentric study.

Authors:  J C Gentric; A Biondi; M Piotin; C Mounayer; K Lobotesis; A Bonafé; V Costalat
Journal:  AJNR Am J Neuroradiol       Date:  2013-01-24       Impact factor: 3.825

View more