Dose perturbations due to contrast medium and air in mammosite treatment: an experimental and Monte Carlo study.

C W Cheng1, R Mitra, X Allen Li, Indra J Das.   

Abstract

In the management of early breast cancer, a partial breast irradiation technique called MammoSite (Proxima Therapeutic Inc., Alpharetta, GA) has been advocated in recent years. In MammoSite, a balloon implanted at the surgical cavity during tumor excision is filled with a radio-opaque solution, and radiation is delivered via a high dose rate brachytherapy source situated at the center of the balloon. Frequently air may be introduced during placement of the balloon and/or injection of the contrast solution into the balloon. The purpose of this work is to quantify as well as to understand dose perturbations due to the presence of a high-Z contrast medium and/or an air bubble with measurements and Monte Carlo calculations. In addition, the measured dose distribution is compared with that obtained from a commercial treatment planning system (Nucletron PLATO system). For a balloon diameter of 42 mm, the dose variation as a function of distance from the balloon surface is measured for various concentrations of a radio-opaque solution (in the range 5%-25% by volume) with a small volume parallel plate ion chamber and a micro-diode detector placed perpendicular to the balloon axis. Monte Carlo simulations are performed to provide a basic understanding of the interaction mechanism and the magnitude of dose perturbation at the interface near balloon surface. Our results show that the radio-opaque concentration produces dose perturbation up to 6%. The dose perturbation occurs mostly within the distances <1 mm from the balloon surface. The Plato system that does not include heterogeneity correction may be sufficient for dose planning at distances > or = 10 mm from the balloon surface for the iodine concentrations used in the MammoSite procedures. The dose enhancement effect near the balloon surface (<1 mm) due to the higher iodine concentration is not correctly predicted by the Plato system. The dose near the balloon surface may be increased by 0.5% per cm3 of air. Monte Carlo simulation suggests that the interface effect (enhanced dose near surface) is primarily due to Compton electrons of short range (<0.5 mm). For more accurate dosimetry in MammoSite delivery, the dose perturbation due to the presence of a radio-opaque contrast medium and air bubbles should be considered in a brachytherapy planning system.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16121583     DOI: 10.1118/1.1943827

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  The impact of uncertainties associated with MammoSite brachytherapy on the dose distribution in the breast.

Authors:  Saleh Bensaleh; Eva Bezak
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

2.  Dose perturbation study in a multichannel breast brachytherapy device.

Authors:  Eric D Slessinger; Rodney Fletcher; Indra J Das
Journal:  J Contemp Brachytherapy       Date:  2011-12-30

Review 3.  Review of clinical brachytherapy uncertainties: analysis guidelines of GEC-ESTRO and the AAPM.

Authors:  Christian Kirisits; Mark J Rivard; Dimos Baltas; Facundo Ballester; Marisol De Brabandere; Rob van der Laarse; Yury Niatsetski; Panagiotis Papagiannis; Taran Paulsen Hellebust; Jose Perez-Calatayud; Kari Tanderup; Jack L M Venselaar; Frank-André Siebert
Journal:  Radiother Oncol       Date:  2013-11-30       Impact factor: 6.280

4.  Comparison of a 3-D multi-group SN particle transport code with Monte Carlo for intracavitary brachytherapy of the cervix uteri.

Authors:  Kent A Gifford; Todd A Wareing; Gregory Failla; John L Horton; Patricia J Eifel; Firas Mourtada
Journal:  J Appl Clin Med Phys       Date:  2009-12-03       Impact factor: 2.102

5.  Dose modification factors for 192Ir high-dose-rate irradiation using Monte Carlo simulation.

Authors:  Bassel Kassas; Firas Mourtada; John L Horton; Richard G Lane; Thomas A Buchholz; Eric A Strom
Journal:  J Appl Clin Med Phys       Date:  2006-08-24       Impact factor: 2.102

6.  Dose modification factor analysis of multilumen balloon brachytherapy applicator with Monte Carlo simulation.

Authors:  David Pearson; Eric A Williams
Journal:  J Appl Clin Med Phys       Date:  2013-05-08       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.