Literature DB >> 16121576

Characterization and correction of temporal artifacts in CT.

L Spies1, R Luhta.   

Abstract

Image artifacts, caused by a temporally delayed response of a computed tomography (CT) detector, were investigated. To study its consequences, a computer model for a standard third generation CT scanner was devised and simulations were carried out. Resulting image artifacts were studied for various voxel-based and mathematical phantoms using three time constants, which characterize the delayed signal responses of different detector configurations. Furthermore, a theory was developed to understand temporal artifacts in reconstructed images. A filter function was derived, which compensates for temporal artifacts. For a given phantom, simulations and theory demonstrate that artifacts scale with the time constant, characteristic for the detector, and the angular speed of the scanner.

Mesh:

Year:  2005        PMID: 16121576     DOI: 10.1118/1.1929147

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Conditional Covariances for the Signal Lag Measurements in Fluoroscopic Imaging.

Authors:  Eunae Lee; Dong Sik Kim
Journal:  Diagnostics (Basel)       Date:  2021-12-31
  1 in total