Quality control and diagnostic reference levels in intraoral dental radiographic facilities.

K Hatziioannou1, E Psarouli, E Papanastassiou, P Bousbouras, H Kodona, O Kimoundri, M Delichas.   

Abstract

OBJECTIVES: A quality control program in intraoral radiographic equipment was conducted aiming to verify compliance with the latest legal requirements and to measure radiation dose in order to contribute to the establishment of appropriate diagnostic reference levels.
METHODS: This study was performed in 191 intraoral radiographic facilities in Northern Greece. The quality control program included visual inspection of the unit, registration of its characteristics, measurement and calculation of several radiological parameters, including entrance surface dose, and overall compliance with legislation requirements.
RESULTS: Most units (61.8%) were found to perform within their corresponding specifications. Exposure timer accuracy was the parameter that the dental units showed the poorest performance. Inadequate beam filtration was found in almost half of the older 50 kVp units. Entrance surface dose showed significant discrepancies, varying from 0.30 mGy to 16.09 mGy with a third quartile value of 3.37 mGy.
CONCLUSIONS: A diagnostic reference level of 3.5 mGy is considered to be a realistic value for all intraoral equipment currently operating in Greece. However, for newly installed units, a value of 2.8 mGy is considered to be more appropriate.

Mesh:

Year:  2005        PMID: 16120881     DOI: 10.1259/dmfr/38802780

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  2 in total

1.  A Novel Phantom and a Dedicated Developed Software for Image Quality Controls in X-Ray Intraoral Devices.

Authors:  Giuseppe Acri; Sergio Gurgone; Claudio Iovane; Marco B Romeo; Daniele Borzelli; Barbara Testagrossa
Journal:  J Biomed Phys Eng       Date:  2021-04-01

2.  Effects of Different Viewing Conditions on Radiographic Interpretation.

Authors:  Mahkameh Moshfeghi; Majid Shahbazian; Soodabeh Sadat Sajadi; Sepideh Sajadi; Hossein Ansari
Journal:  J Dent (Tehran)       Date:  2015-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.