Bursting tension of medium and large canine arteries sealed with ultrasonic energy or suture ligation.

Loretta J Bubenik1, Giselle Hosgood, Sunil C Vasanjee.   

Abstract

OBJECTIVE: To compare bursting tension of arteries > or =3 mm in diameter sealed with ultrasonic energy (UE) with arteries ligated with suture. STUDY
DESIGN: Experimental, nested factorial design. SAMPLE POPULATION: Vascular segments from canine cadavers.
METHODS: Arterial segments (12) were collected from each of 16 canine cadavers and equally divided into 2 groups based on vessel diameter (medium, 3.0- < 4.5 mm; large, 4.5 - < or =6.0 mm). Arterial specimens (3) from each group were sealed with either UE (Harmonic Scalpel, HS, power level 3) or suture ligation. The mean bursting pressure and tension were determined and compared.
RESULTS: Bursting tension was significantly higher (P< .0001) for sutured arteries than UE-occluded arteries, irrespective of size. Bursting tension was significantly higher (P=.0013) for medium than large UE-occluded arteries, whereas there was no difference associated with size for ligated arteries. UE energy failed to seal 1 medium artery and 7 large arteries. Compared with normal blood pressure, bursting pressures were 3.5 times greater for sutured arteries irrespective of size, 1.8 times greater for medium UE-occluded arteries, and approached normal blood pressure for large UE-occluded arteries.
CONCLUSION: At the power level tested, UE should not be used to seal arteries > 4.5 mm in diameter. Suture provided an optimal seal for arteries < or =6.0 mm in diameter. CLINICAL RELEVANCE: Use of UE to occlude arteries during laparoscopic surgery is advantageous because only 1 instrument is required to simultaneously cut and coagulate tissue, but care should be exercised where large arteries might be encountered.

Entities:  

Mesh:

Year:  2005        PMID: 16115087     DOI: 10.1111/j.1532-950x.2005.00043.x

Source DB:  PubMed          Journal:  Vet Surg        ISSN: 0161-3499            Impact factor:   1.495


  6 in total

1.  Comparison of four energy-based vascular sealing and cutting instruments: A porcine model.

Authors:  Benjamin Person; David A Vivas; Dan Ruiz; Michael Talcott; James E Coad; Steven D Wexner
Journal:  Surg Endosc       Date:  2007-12-20       Impact factor: 4.584

2.  Pulse-wave propagation in straight-geometry vessels for stiffness estimation: theory, simulations, phantoms and in vitro findings.

Authors:  Danial Shahmirzadi; Ronny X Li; Elisa E Konofagou
Journal:  J Biomech Eng       Date:  2012-11       Impact factor: 2.097

3.  Efficacy and safety of 5-mm-diameter bipolar and ultrasonic shears for cutting carotid arteries of the hybrid pig.

Authors:  René Mantke; W Halangk; A Habermann; B Peters; S Konrad; M Guenther; H Lippert
Journal:  Surg Endosc       Date:  2010-07-08       Impact factor: 4.584

4.  Design of a factorial experiment with randomization restrictions to assess medical device performance on vascular tissue.

Authors:  Wiebke S Diestelkamp; Carissa M Krane; Margaret F Pinnell
Journal:  BMC Med Res Methodol       Date:  2011-05-20       Impact factor: 4.615

5.  Postoperative efficacy and safety of vessel sealing: an experimental study on carotid arteries of the pig.

Authors:  Stéphane V Berdah; Christiaan Hoff; Peiman Hossein Poornoroozy; Peter Razek; Yves Van Nieuwenhove
Journal:  Surg Endosc       Date:  2012-02-21       Impact factor: 4.584

6.  Skeletonization and Isolation of the Glissonean and Venous Branches in Liver Surgery With an Ultrasonic Scalpel Technology.

Authors:  Takeshi Aoki; Masahiko Murakami; Tomotake Koizumi; Yuta Enami; Reiko Koike; Akira Fujimori; Tomokazu Kusano; Kazuhiro Matsuda; Kosuke Yamada; Koji Nogaki; Makoto Watanabe; Koji Otsuka; Haytham Gareer; Takashi Kato
Journal:  Int Surg       Date:  2015-06
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.