[Accuracy, utility and complications of continuous glucose monitoring system (CGMS) in pediatric patients with type 1 diabetes].

Frederico F R Maia1, Levimar R Araújo.   

Abstract

OBJECTIVE: To evaluate the accuracy, utility and complications of continuous glucose monitoring system in children and adolescents with type 1 diabetes.
METHODS: This retrospective study assessed 16 type 1 diabetic patients (16.12+/-4.41 years) submitted to continuous glucose monitoring system (Medtronic; Northridge, CA) for 72 hours. The following parameters were analyzed: mean capillary glucose level and mean glucose value measured by the continuous glucose monitoring system; glucose excursions (continuous glucose monitoring system vs. capillary glucose measurement), postprandial hyperglycemia (NR < 140 mg/dl), nocturnal hypoglycemia, complications (trauma, local infection, disconnection) and therapeutic management after continuous glucose monitoring. A1c levels were measured at the beginning and after 3 months of the study.
RESULTS: The mean capillary glucose values were 214.3+/-66.5 mg/dl vs. 207.6+/-54.6 mg/dl by continuous glucose monitoring system, with a significant correlation (p = 0.001). The correlation coefficient and mean absolute error were 0.86+/-0.21 and 12.6% of the median, respectively. The continuous glucose monitoring system was significantly more efficient in detecting glucose excursion than fingerstick capillary blood sampling (p = 0.04; W = 74), and postprandial hyperglycemia was identified in 60% of type 1 diabetic patients with a median value of 157 mg/dl (< 140 mg/dl). Nocturnal hypoglycemia was detected in 46.7% of these patients. The evaluation of A1c levels in eight (50%) patients before continuous glucose monitoring and after 3 months showed a significantly lower level of A1c in this population (8.18+/-1.5 vs. 7.28+/-1.3; p = 0.034). The therapeutic management of type 1 diabetes was changed in 100% of patients. No complications were detected in 93.7% of patients.
CONCLUSIONS: The continuous glucose monitoring system showed to be a very safe, well-tolerated and highly accurate method, with a low complication rate. It is a good method to identify glucose excursion and postprandial hyperglycemia, and to improve metabolic changes in therapeutic strategies, with a significant impact on the A1c levels of pediatric diabetic patients. The efficacy of the continuous glucose monitoring system in detecting hypoglycemia is still unclear in the medical literature.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16106313

Source DB:  PubMed          Journal:  J Pediatr (Rio J)        ISSN: 0021-7557            Impact factor:   2.197


  3 in total

1.  Effect of short-term use of a continuous glucose monitoring system with a real-time glucose display and a low glucose alarm on incidence and duration of hypoglycemia in a home setting in type 1 diabetes mellitus.

Authors:  Raymond J Davey; Timothy W Jones; Paul A Fournier
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

2.  Real-time continuous glucose monitoring in the clinical setting: the good, the bad, and the practical.

Authors:  Irene Mamkin; Svetlana Ten; Sonal Bhandari; Neesha Ramchandani
Journal:  J Diabetes Sci Technol       Date:  2008-09

3.  New technologies for diabetes: a review of the present and the future.

Authors:  Neesha Ramchandani; Rubina A Heptulla
Journal:  Int J Pediatr Endocrinol       Date:  2012-10-26
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.