Abnormal responses of electronic pocket dosimeters caused by high frequency electromagnetic fields emitted from digital cellular telephones.

Shizuhiko Deji1, Kunihide Nishizawa.   

Abstract

High frequency electromagnetic fields emitted from digital cellular telephones (cell phones) occasionally cause abnormally high and erroneous indicated dose readings on electronic pocket dosimeters (EPDs). Electric field strength distribution around a cell phone transmitting 1.5 GHz band with a maximum power of 0.8 W was analyzed by using an isotropic probe with tri-axial dipole antennas. Five types of EPDs were exposed to the fields for 50 s under configurations relative to the cell phone. The electric field distribution expanded around the phone's antenna and had a maximum electric field strength of 36.5 +/- 0.3 V m(-1). The cell phone gave rise to erroneous indicated dose readings on four out of five EPDs. The maximum value of erroneous indicated dosage for 50 s reached 1,283 microSv, which was about 2.6% of the annual effective dose limit of 50 mSv. The electromagnetic susceptibility of the EPDs was higher in the sections where the semiconductor detectors or electric circuit boards were located. The distance required to prevent electromagnetic interference differed for each EPD and ranged from 2.0 to 21.0 cm from the cell phone. The electric and magnetic field immunity levels of the EPDs varied from 9.2 V m(-1) to greater than 37.6 V m(-1), and from 0.03 A m(-1) to greater than 0.51 A m(-1). The EPDs displayed erroneous dose readings during exposure but recovered their normal performance after the cell phone ceased transmitting. The electromagnetic immunity levels of the EPDs were either equal to or greater than the IEC-standard. The immunity levels should be enhanced greater than the IEC-standard from the standpoint of radiation protection. The simplest and most reliable measure to prevent potential malfunction is to prohibit the radiation workers from carrying cell phones to their workplace.

Entities:  

Mesh:

Year:  2005        PMID: 16096498     DOI: 10.1097/01.hp.0000164652.70851.4f

Source DB:  PubMed          Journal:  Health Phys        ISSN: 0017-9078            Impact factor:   1.316


  1 in total

1.  Neural network model applied to electromagnetic shielding effectiveness of ultra-light Ni/Cu coated polyester fibrous materials.

Authors:  Aravin Prince Periyasamy; Lekha Priya Muthusamy; Jiri Militký
Journal:  Sci Rep       Date:  2022-05-21       Impact factor: 4.996

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.