Literature DB >> 16093625

Accuracy of MR temperature measurement based on chemical shift change for radiofrequency ablation using hook-shaped electrodes.

Shigeo Okuda1, Kagayaki Kuroda, Osamu Kainuma, Koichi Oshio, Hirokazu Fujiwara, Sachio Kuribayashi.   

Abstract

The purpose of the current study is to evaluate the accuracy of MR thermometry for radiofrequency ablation (RFA) with hook-shaped electrodes. The objects were eight extracted bovine livers. The chemical shift change was calculated from MR images acquired with a spoiled gradient echo sequence and compared with the temperature directly measured with a thermocouple. Linear regression was established between them with a coefficient of -0.0110+/-0.0007 ppm/ degrees C and errors were calculated as -0.50+/-7.50 degrees C. MR thermometry is capable of monitoring temperature for RFA.

Entities:  

Mesh:

Year:  2004        PMID: 16093625     DOI: 10.2463/mrms.3.95

Source DB:  PubMed          Journal:  Magn Reson Med Sci        ISSN: 1347-3182            Impact factor:   2.471


  2 in total

1.  The influence of Nd:YAG laser irradiation on Fluoroptic® temperature measurement: an experimental evaluation.

Authors:  Frank Hübner; Babak Bazrafshan; Jörg Roland; Antje Kickhefel; Thomas Josef Vogl
Journal:  Lasers Med Sci       Date:  2012-04-04       Impact factor: 3.161

Review 2.  New horizons in MR-controlled and monitored radiofrequency ablation of liver tumours.

Authors:  Alexandru Cernicanu; Matthieu Lepetit-Coiffé; Magalie Viallon; Sylvain Terraz; Christoph D Becker
Journal:  Cancer Imaging       Date:  2007-11-05       Impact factor: 3.909

  2 in total