Shigeo Okuda1, Kagayaki Kuroda, Osamu Kainuma, Koichi Oshio, Hirokazu Fujiwara, Sachio Kuribayashi.
Abstract
The purpose of the current study is to evaluate the accuracy of MR thermometry for radiofrequency ablation (RFA) with hook-shaped electrodes. The objects were eight extracted bovine livers. The chemical shift change was calculated from MR images acquired with a spoiled gradient echo sequence and compared with the temperature directly measured with a thermocouple. Linear regression was established between them with a coefficient of -0.0110+/-0.0007 ppm/ degrees C and errors were calculated as -0.50+/-7.50 degrees C. MR thermometry is capable of monitoring temperature for RFA.Entities:
Mesh:
Year: 2004 PMID: 16093625 DOI: 10.2463/mrms.3.95
Source DB: PubMed Journal: Magn Reson Med Sci ISSN: 1347-3182 Impact factor: 2.471