Output factor measurements for a kilovoltage X-ray therapy unit.

B J Healy1, A Gibbs, R L Murry, J E Prunster, K N Nitschke.   

Abstract

Output factors at the surface for treatment cones and lead cut-outs have been measured for a Pantak Therapax SXT 150 superficial therapy unit with x-ray beam qualities from 1 to 13 mm A1 HVL. A variety of phantom materials and two ionisation chambers were tested for their suitability in output factor and percentage depth dose measurement. Solid water proved a useful water-equivalent phantom material with discrepancies between measurements in water and solid water less than 2.3% for percentage depth dose and less than 0.6% for output factors. Larger measurement discrepancies were found for Plastic Water and Perspex. A PTW Markus chamber was found to compare well with a NE 2532/3 low energy chamber in percentage depth dose measurement, but discrepancies arose between the chambers in output factor measurements, up to 5% for small field sizes. Measurements indicated that the Markus chamber had an energy dependent response in the kilovoltage range, which could account for the discrepancy in output factor measurement.

Entities:  

Mesh:

Year:  2005        PMID: 16060318     DOI: 10.1007/bf03178702

Source DB:  PubMed          Journal:  Australas Phys Eng Sci Med        ISSN: 0158-9938            Impact factor:   1.430


  2 in total

1.  Use of calculations to validate beam quality and relative dose measurements for a kilovoltage X-ray therapy unit.

Authors:  B J Healy; R F Hill
Journal:  Phys Eng Sci Med       Date:  2022-04-05

2.  Comparison of an anthropomorphic PRESAGE® dosimeter and radiochromic film with a commercial radiation treatment planning system for breast IMRT: a feasibility study.

Authors:  Khalid Iqbal; Kent A Gifford; Geoffrey Ibbott; Ryan L Grant; Saeed Buzdar
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.