Measurement of serum salicylate levels by solid-phase extraction and desorption/ionization on silicon mass spectrometry.

Shoji Okuno1, Yoshinao Wada.   

Abstract

The applicability of the matrix-free laser desorption/ionization on silicon mass spectrometry (DIOS-MS) to measuring serum drug levels was examined by analyzing serum salicylic acid. The optimized and simple solid-phase extraction (SPE) allowed good recovery, 88.9 +/- 5.8%, for 1.4 mM (200 mg/L) of salicylic acid in serum. The negative ion MS allowed measurements of deprotonated molecules without interference from other signals. Using a deuterium-labeled internal standard, good linearity was obtained in the 0.14 to 4.2 mM (20-600 mg/L) range, which was sufficient for monitoring the therapeutic anti-inflammatory dose. SPE followed by DIOS-MS is anticipated to be a method of measuring drug levels in blood and may allow high throughput analysis.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16059881     DOI: 10.1002/jms.859

Source DB:  PubMed          Journal:  J Mass Spectrom        ISSN: 1076-5174            Impact factor:   1.982