Literature DB >> 1605952

Common artifacts encountered in thoracic magnetic resonance imaging: recognition, derivation, and solutions.

M L Schiebler1, J Listerud.   

Abstract

The article discusses the many types of thoracic magnetic resonance imaging artifacts, their derivations, and their current solutions. Magnetic resonance imaging artifacts of the thorax can be divided into two major categories. First, there are machine-related artifacts related to the machine's hardware: the intrinsic magnetic field, chemical shift, magnetic susceptibility, radio frequency leaks, metal artifacts, B1-homogeneity, gradient coils, truncation, aliasing, zipper artifacts, artifacts related to the surface coil profiles, pulse profiles, and crosstalk. Second, there are artifacts related to motion: voluntary patient motion, involuntary motion, and physiologic motions. Each one of these artifacts, and their solutions, will be discussed.

Entities:  

Mesh:

Year:  1992        PMID: 1605952

Source DB:  PubMed          Journal:  Top Magn Reson Imaging        ISSN: 0899-3459


  2 in total

Review 1.  Artifacts in body MR imaging: their appearance and how to eliminate them.

Authors:  Alfred Stadler; Wolfgang Schima; Ahmed Ba-Ssalamah; Joachim Kettenbach; Edith Eisenhuber
Journal:  Eur Radiol       Date:  2006-12-06       Impact factor: 5.315

Review 2.  Current artefacts in cardiac and chest magnetic resonance imaging: tips and tricks.

Authors:  Khalid Alfudhili; Pier G Masci; Jean Delacoste; Jean-B Ledoux; Grégoire Berchier; Vincent Dunet; Salah D Qanadli; Juerg Schwitter; Catherine Beigelman-Aubry
Journal:  Br J Radiol       Date:  2016-03-17       Impact factor: 3.039

  2 in total