Literature DB >> 16056078

Pullout strength of self-tapping screws inserted to different depths.

Robert Berkowitz1, Glen Njus, Gregory Vrabec.   

Abstract

OBJECTIVES: The purpose of this study was to determine whether the depth of insertion through the far cortex of self-tapping screws significantly affects pullout strength.
DESIGN: Fifty, Synthes, 3.5-mm, self-tapping screws were inserted into synthetic bone blocks and divided into 5 groups. Group 1 had screws with their tips inserted 1 mm short of the far cortex. Group 2 had screws inserted flush with the far cortex. Groups 3, 4, and 5 had screws inserted 1 mm, 2 mm, and 3 mm past the far cortex respectively. Pullout strength was then tested.
SETTING: Institutional research laboratory. MAIN OUTCOME MEASUREMENTS: Pullout strength (peak force) was measured for each group and analyzed using a single factor analysis of variance-balanced incomplete block design.
RESULTS: Peak force values presented as mean +/- SD for the 5 groups were as follows: group 1 (1380 +/- 69 N), group 2 (1566 +/- 137 N), group 3 (1956 +/- 137 N), group 4 (2013 +/- 184 N), group 5 (2044 +/- 174 N). With a P < or = 0.05, it was found that groups I and II had statistically different pullout strengths than all other groups. However, there was no significant difference in pullout strength between groups 3, 4, and 5.
CONCLUSIONS: Synthes self-tapping screws exhibit their highest pullout strength when inserted 1 mm past the far cortex, and there is no significant increase in pullout strength with deeper insertion depths.

Entities:  

Mesh:

Year:  2005        PMID: 16056078     DOI: 10.1097/01.bot.0000161544.72757.95

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  3 in total

1.  Pullout Strength After Multiple Reinsertions in Radial Bone Fixation.

Authors:  Alfonso Mejia; Giovanni Solitro; Elena Gonzalez; Amit Parekh; Mark Gonzalez; Farid Amirouche
Journal:  Hand (N Y)       Date:  2018-09-06

2.  Factors Associated With the Accuracy of Depth Gauge Measurements.

Authors:  Pengcheng Liu; Joanna Xi Xiao; Chen Zhao; Xiaodong Li; Guantong Sun; Fei Yang; Xiaoqing Wang
Journal:  Front Surg       Date:  2022-01-13

3.  Evaluation of the effects of different surface configurations on stability of miniscrews.

Authors:  Tolga Topcuoglu; A Altuğ Bicakci; M Cihat Avunduk; Z Deniz Sahin Inan
Journal:  ScientificWorldJournal       Date:  2013-07-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.