[Experimental and theoretical studies on radiation protective effect of a lighter non-lead protective apron].

Yoshihisa Takano1, Keiichiro Okazaki, Koji Ono, Michiaki Kai.   

Abstract

Non-lead aprons using composite materials are often used for radiation protective aprons instead of heavy lead aprons. However, the protective effect of the lighter, non-lead aprons has not been well evaluated, and it is not yet clear how they compare with lead aprons. Therefore, we investigated the protective performance of non-lead aprons theoretically and experimentally by comparing them with lead aprons under clinical conditions. We measured the energy spectra for direct and scattered X-rays passing through protective aprons or not, and measured doses with glass dosimeters for validation of theoretical calculations based on the energy spectra. We found that the protective effect of non-lead aprons was higher than that of lead aprons at X-ray of tube voltages of 70-100 kV, which are often used for radiography and fluorography. This demonstrated that the non-lead aprons are more useful in many situations than heavy lead aprons.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16049417     DOI: 10.6009/jjrt.kj00003326912

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  1 in total

1.  Evaluation of the effectiveness of X-ray protective aprons in experimental and practical fields.

Authors:  Hiroshige Mori; Kichiro Koshida; Osamu Ishigamori; Kosuke Matsubara