The effect of Butterworth and Metz reconstruction filters on volume and ejection fraction calculations with 99Tcm gated myocardial SPECT.

T Vakhtangandze1, D O Hall, F V Zananiri, M R Rees.   

Abstract

This study was carried out to measure the differences produced by change of reconstruction filter in calculations of left-ventricular end-diastolic volumes, end-systolic volumes, stroke-volumes and left-ventricular ejection-fractions from (99)Tc(m) Sestamibi (Bristol-Myers Squibb) gated myocardial perfusion SPECT studies. 30 patients had gated SPECT myocardial perfusion imaging at rest. The acquired projections were separately filtered with two filters, a low-pass filter (Butterworth) and an edge-enhancement filter (Metz). Each study was then further processed to determine left-ventricular end-diastolic volume, end-systolic volume, stroke volume and ejection fraction, and to assess defect size. The results for each patient with the two filters were compared. Calculated end-diastolic volumes, end-systolic volumes and left-ventricular ejection fractions, for each filter, were well correlated. Stroke volumes showed worse correlation. The differences between left-ventricular ejection-fractions, end-diastolic volumes and end-systolic volumes were statistically significant. There was no significant difference in stroke volumes. Ejection fractions were inversely correlated with defect size, but change in ejection fraction due to filter was not. End-diastolic and end-systolic volumes were correlated with defect size, but change in volumes due to filter was not. Thus the results for changes produced by choice of filter are not dependent on defect size. Using different reconstruction pre-filters in gated myocardial perfusion SPECT significantly changes the results of calculations of physiological parameters. Each centre should be consistent in the use of filters as this may affect the clinical consequences of the result.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16046425     DOI: 10.1259/bjr/25662922

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  4 in total

1.  Pix2Pix generative adversarial network for low dose myocardial perfusion SPECT denoising.

Authors:  Jingzhang Sun; Yu Du; ChienYing Li; Tung-Hsin Wu; BangHung Yang; Greta S P Mok
Journal:  Quant Imaging Med Surg       Date:  2022-07

2.  Do reconstruction filters really effect the volume and ejection fraction calculation with 99Tc-sestamibigated myocardial SPECT?

Authors:  Ora Manish; Subhash Chand Kheruka; Sukanta Barai; Sanjay Gambhir
Journal:  Indian J Nucl Med       Date:  2010-10

3.  Late gadolinium uptake demonstrated with magnetic resonance in patients where automated PERFIT analysis of myocardial SPECT suggests irreversible perfusion defect.

Authors:  Lene Rosendahl; Peter Blomstrand; Jan L Ohlsson; Per-Gunnar Björklund; Britt-Marie Ahlander; Sven-Ake Starck; Jan E Engvall
Journal:  BMC Med Imaging       Date:  2008-12-12       Impact factor: 1.930

Review 4.  Filters in 2D and 3D Cardiac SPECT Image Processing.

Authors:  Maria Lyra; Agapi Ploussi; Maritina Rouchota; Stella Synefia
Journal:  Cardiol Res Pract       Date:  2014-04-01       Impact factor: 1.866

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.