N Lindegårdh1, T Singtoroj, A Annerberg, N J White, N P J Day.
Abstract
A bioanalytic method for the determination of amoxicillin in plasma by hydrophilic interaction solid-phase extraction and liquid chromatography has been developed and validated. Plasma was precipitated with acetonitrile before samples were loaded onto a zwitterionic hydrophilic interaction liquid chromatography (ZIC-HILIC) solid-phase extraction column. Amoxicillin was analyzed by liquid chromatography on an Aquasil (150 x 4.6 mm) LC column with mobile-phase acetonitrile: phosphate buffer (pH 2.5; 0.1 mol/L) (7:93, v/v) and UV detection at 230 nm. A regression model using 1/concentration weighting was found the most appropriate for quantification. The intraassay precision for plasma was 3.3% at 15.0 microg/mL and 10.9% at 0.200 microg/mL. The interassay precision for plasma was 1.8% at 15.0 microg/mL and 7.5% at 0.200 microg/mL. The total-assay precision for plasma over 4 days using a total of 20 replicates was 13.2%, 5.5%, and 3.8% at 0.200 microg/mL, 3.00 microg/mL, and 15.0 microg/mL, respectively. The lower limit of quantification and the limit of detection were 0.050 microg/mL and 0.025 microg/mL, respectively, for 100 microL plasma. Long-term storage stability studies of amoxicillin in plasma indicate that a temperature of -80 degrees C is necessary to prevent degradation of amoxicillin.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16044109 DOI: 10.1097/01.ftd.0000158082.38330.85
Source DB: PubMed Journal: Ther Drug Monit ISSN: 0163-4356 Impact factor: 3.681