Literature DB >> 16044102

Hydroxyzine and metabolites as a source of interference in carbamazepine particle-enhanced turbidimetric inhibition immunoassay (PETINIA).

François Parant1, Mustapha Moulsma, Marie Claude Gagnieu, Gisèle Lardet.   

Abstract

A 32-year-old epileptic patient with a lengthy history of multiple-drug abuse and psychotic disorders was found to have an elevated serum carbamazepine concentration of 40.5 mg/L (therapeutic range 4-12 mg/L) using particle-enhanced turbidimetric inhibition immunoassay (PETINIA). Serum reanalysis by LC-DAD revealed only high hydroxyzine (HDZ) concentration (HDZ = 0.55 mg/L; therapeutic range <0.1 mg/L), suggesting cross-reactivity between HDZ and PETINIA. To confirm this hypothesis, the authors tested 2 commercially available carbamazepine immunoassays, PETINIA and EMIT 2000, for in vitro cross-reactivity with HDZ and 2 HDZ metabolites (cetirizine and norchlorcyclizine). To determine the frequency of this interaction in a clinical setting, 40 sera of 39 patients taking HDZ without carbamazepine were tested by both immunoassays. For some samples, LC-ESI-MS analysis of HDZ metabolites was performed. Additionally, cross-reactivities produced by other benzhydrylpiperazine drugs were evaluated. in vitro, 5 mg of HDZ, cetirizine, and norchlorcyclizine cross-reacted with PETINIA at 85%, 125%, and 66%, respectively. Conversely, EMIT 2000 showed no cross-reactivity. For PETINIA, erroneous carbamazepine concentrations were detected in 35 out of 40 sera of patients taking HDZ. The magnitude of interference correlated moderately with serum HDZ concentrations (Spearman rho coefficient 0.58, P < 0.001), suggesting a major role for the multiple HDZ metabolites (4 serum metabolites were detected by LC-ESI-MS). Furthermore, other benzhydrylpiperazine drugs (eg, oxatomide) showed in vitro cross-reactivity with PETINIA. In conclusion, HDZ and its metabolites cross-react with carbamazepine PETINIA immunoassay, which could significantly affect the correct interpretation of serum carbamazepine concentrations in patients treated with HDZ.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16044102     DOI: 10.1097/01.ftd.0000160718.88822.20

Source DB:  PubMed          Journal:  Ther Drug Monit        ISSN: 0163-4356            Impact factor:   3.681


  2 in total

1.  Turbidimetric carbamazepine immunoassay on the ADVIA 1650 and 2400 analyzers is free from interference of antihistamine drugs hydroxyzine and cetirizine.

Authors:  Pradip Datta; Amitava Dasgupta
Journal:  J Clin Lab Anal       Date:  2007       Impact factor: 2.352

2.  Analytical performance evaluation of ADVIA Chemistry Carbamazepine_2 assay: minimal cross-reactivity with carbamazepine 10, 11-epoxide and none with hydroxyzine or cetirizine.

Authors:  Amitava Dasgupta; Meredith A Reyes; Barbara G Davis; Anne M Marlow; Myrtle Johnson
Journal:  J Clin Lab Anal       Date:  2010       Impact factor: 2.352

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.