Literature DB >> 16039503

Experimental evaluation of the Corneal Concept 360 intraocular lens with the Miyake-Apple view.

Liliana Werner1, M Scott Hickman, Russell M LeBoyer, Nick Mamalis.   

Abstract

PURPOSE: To evaluate a new single-piece, hydrophilic acrylic intraocular lens (IOL) designed to keep the anterior capsule away from the IOL optic, create the effect of a capsular tension ring, and prevent posterior capsule opacification (PCO).
SETTING: John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.
MATERIALS AND METHODS: Four Concept 360 IOLs (Corneal Laboratoire) were implanted in 4 human cadaver eyes of various sizes (Miyake-Apple technique). Lens centration, ovalization of the capsular bag and capsulorhexis opening, distance between the anterior optic surface and the anterior capsule and between the posterior optic surface and the posterior capsule, as well as presence of capsular striae were evaluated. The capsular bag was stained with trypan blue, and the ophthalmic viscosurgical device (OVD) was stained with fluorescein to enhance visualization.
RESULTS: In each eye, the lens conformed to the capsular bag with precise centration. No ovalization or distortion of the capsular bag or the capsulorhexis and no posterior capsule striae were observed. Measurements of the capsulorhexis diameter before and after IOL implantation were the same. The anterior capsule remained at distance from the anterior optic surface for 360 degrees, and tight contact between the posterior optic edge and the posterior capsule was observed.
CONCLUSIONS: Use of the Miyake-Apple technique is useful in the evaluation of new IOL designs in experimental studies. Design features of the Concept 360 IOL are likely to decrease the outcome of postoperative capsular bag opacification. Long-term clinical studies are necessary to confirm these findings.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16039503     DOI: 10.1016/j.jcrs.2004.10.072

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

1.  Lenticular Imaging: A New Experimental and Quantitative Analysis of Capsular Dynamics, "Choi-Apple View".

Authors:  Hyeck Soo Son; Ko Un Shin; Donghan Kim; Patrick Merz; Tamer Tandogan; Ramin Khoramnia; Gerd U Auffarth; Elfriede Friedmann; Suehey Chae; Kyung Heon Lee; Chul Young Choi
Journal:  Transl Vis Sci Technol       Date:  2019-05-20       Impact factor: 3.283

2.  Post-operative capsular opacification: a review.

Authors:  Shetal M Raj; Abhay R Vasavada; S R Kaid Johar; Vaishali A Vasavada; Viraj A Vasavada
Journal:  Int J Biomed Sci       Date:  2007-12
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.