Literature DB >> 16039488

Clinical outcomes of phakic refractive lens in myopic and hyperopic eyes: 1-year results.

Annemari Koivula1, Anders Petrelius, Charlotta Zetterström.   

Abstract

PURPOSE: To confirm the safety, efficacy, and predictability of the surgical correction of myopia and hyperopia with the phakic refractive lens (PRL) (Medennium Inc.).
SETTING: St. Eriks Eye Hospital, Stockholm, Sweden.
METHODS: This was a prospective clinical study of 20 eyes, 14 myopic and 6 hyperopic, that had PRL implantation at St. Eriks Eye Hospital from April to November 2002. Examinations were performed preoperatively and 1 day, 1 week, 3 months, and 1 year postoperatively. Follow-up included evaluation of the PRL rotation with retroillumination photography, evaluation of the distance between the PRL and the crystalline lens with Scheimpflug image, laser flare, endothelial cell count, uncorrected (UCVA) and best corrected (BCVA) visual acuity, residual refractive error, refractive stability, intraocular pressure, and induced cataract.
RESULTS: Postoperatively, 11 eyes (55%) gained 1 or more lines, 5 eyes (25%) had no change, and 4 eyes (20%) lost 1 line of BCVA. No eye lost 2 or more lines. Mean UCVA was 0.87+/- 0.29 postoperatively. Laser flare values were highest 1 day after operation with normalization at 3 months and without changes at 1 year (P<.05). A rotation of 10 degrees or more was found in 15 eyes (75%) during the first year. The distance between the PRL and crystalline lens was considerably less at 1 year than at baseline (P<.05). There was no statistically significant endothelial cell loss induced by the PRL (P<.05). No induced cataract, glaucoma, or inflammation was observed. In 1 hyperopic eye, horizontal iris transillumination defects were noticed at 1 year.
CONCLUSION: Safety and efficacy indexes were high at 1-year follow-up. The PRL rotated slightly in the posterior chamber. The distance between the PRL and the crystalline lens was considerably less at 1 year than at baseline.

Entities:  

Mesh:

Year:  2005        PMID: 16039488     DOI: 10.1016/j.jcrs.2004.11.059

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

Review 1.  The posterior chamber phakic refractive lens (PRL): a review.

Authors:  R J Pérez-Cambrodí; D P Piñero; T Ferrer-Blasco; A Cerviño; R Brautaset
Journal:  Eye (Lond)       Date:  2012-12-07       Impact factor: 3.775

2.  Traumatic dislocation of implanted collamer phakic lens: a case report and review of the literature.

Authors:  Majid Moshirfar; Brian C Stagg; Valliammai Muthappan; Shail A Vasavada
Journal:  Open Ophthalmol J       Date:  2014-05-30

Review 3.  Phakic intraocular lenses: An overview.

Authors:  Soraya M R Jonker; Tos T J M Berendschot; Isabelle E Y Saelens; Noël J C Bauer; Rudy M M A Nuijts
Journal:  Indian J Ophthalmol       Date:  2020-12       Impact factor: 1.848

  3 in total