Literature DB >> 16038390

A potential propensity for failure secondary to clot embolism in neonatal ECMO.

Espeed Khoshbin1, David Machin, Hilliary Killer, Giles J Peek, Andrzej W Sosnowski, Richard K Firmin.   

Abstract

OBJECTIVE: To report a single case of oxygenator failure caused by clot embolism originating from the bladder; and to discuss some preventative options. CASE REPORT: A 2.5 kg neonate with a diagnosis of influenza A received veno-arterial (V-A) extracorporeal membrane oxygenation (ECMO) for cardiorespiratory support. Halfway through treatment, she underwent an elective circuit change for numerous clots in her circuit. The patient continued to consume vast quantities of platelets and developed a fatal oxygenator failure after 18 days. DISCUSSION: Amongst the factors influencing the outcome in events of a sudden unexpected oxygenator failure are the severity of patient illness, the size of the clot relative to the size of the oxygenator, the availability of a previously primed circuit and the ease and speed of priming a new oxygenator.
CONCLUSION: There is a need for improvement in the design of small oxygenators and ECMO circuits. Adjustment of the coagulation parameters and lowering the tolerance towards clots in the circuit by electively changing them may reduce the incidence of sudden unexpected oxygenator failure. However, using a slightly larger Medos oxygenator may gain valuable time needed to arrange an oxygenator/circuit change.

Entities:  

Mesh:

Year:  2005        PMID: 16038390     DOI: 10.1191/0267659105pf798cs

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  1 in total

1.  A description of a prototype miniature extracorporeal membrane oxygenation circuit using current technologies in a sheep model.

Authors:  Bryan Terry; Gordy Gunst; Richard Melchior; David Wolfe; Nancy Feocco; Susan Graham; Bruce Searles; Edward Darling
Journal:  J Extra Corpor Technol       Date:  2005-09
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.