Literature DB >> 16028248

Co-registration of x-ray and MR fields of view in a hybrid XMR system.

Huanzhou Yu1, Rebecca Fahrig, Norbert J Pelc.   

Abstract

PURPOSE: To validate one possible function of a real-time x-ray/MR (XMR) interface in a hybrid XMR system using x-ray images as "scouts" to prescribe the MR slices.
MATERIALS AND METHODS: The registration process consists of two steps: 1) calibration, in which the system's geometric parameters are found from fiducial-based registration; and 2) application, in which the x-ray image of a target structure and the estimated geometric parameters are used to prescribe an MR slice to observe the target structure. Errors from the noise in the location of the fiducial markers, and MR gradient nonlinearity were studied. Computer simulations were used to provide guidelines for fiducial marker placement and tolerable error estimation. A least-squares-based correction method was developed to reduce errors from gradient nonlinearity.
RESULTS: In simulations with both sources of errors and the correction for gradient nonlinearity, the use of 16 fiducial markers yielded a mean error of about 0.4 mm over a 7200 cm(3) volume. Phantom scans showed that the prescribed target slice hit most of the target line, and that the length visualized was improved with the least-squares correction.
CONCLUSION: The use of 16 fiducial markers to co-register XMR FOVs can offer satisfactory accuracy in both simulations and experiments. (c) 2005 Wiley-Liss, Inc.

Mesh:

Year:  2005        PMID: 16028248     DOI: 10.1002/jmri.20376

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  4 in total

1.  High-resolution 3D arteriography of chronic total peripheral occlusions using a T1-W turbo spin-echo sequence with inner-volume imaging.

Authors:  Smita Sampath; Amish N Raval; Robert J Lederman; Elliot R McVeigh
Journal:  Magn Reson Med       Date:  2007-01       Impact factor: 4.668

2.  Robust automatic rigid registration of MRI and X-ray using external fiducial markers for XFM-guided interventional procedures.

Authors:  Ashvin K George; Merdim Sonmez; Robert J Lederman; Anthony Z Faranesh
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

3.  Novel System for Real-Time Integration of 3-D Echocardiography and Fluoroscopy for Image-Guided Cardiac Interventions: Preclinical Validation and Clinical Feasibility Evaluation.

Authors:  Aruna V Arujuna; R James Housden; Yingliang Ma; Ronak Rajani; Gang Gao; Niels Nijhof; Pascal Cathier; Roland Bullens; Geert Gijsbers; Victoria Parish; Stamatis Kapetanakis; Jane Hancock; C Aldo Rinaldi; Michael Cooklin; Jaswinder Gill; Martyn Thomas; Mark D O'neill; Reza Razavi; Kawal S Rhode
Journal:  IEEE J Transl Eng Health Med       Date:  2014-01-30       Impact factor: 3.316

4.  Ultrasound fusion imaging with real-time navigation for the surveillance after endovascular aortic aneurysm repair.

Authors:  Jin Hyun Joh; Sang-Ah Han; Sang-Hyun Kim; Ho-Chul Park
Journal:  Ann Surg Treat Res       Date:  2017-05-29       Impact factor: 1.859