PURPOSE: Bioabsorbable interference screw design may influence biomechanical characteristics. This study compared the insertion torque and load at failure characteristics of 2 types of screws during retrograde fixation of a soft-tissue graft. TYPE OF STUDY: Biomechanical study.
METHODS: Eight matched pairs of doubled 100-mm long tibialis anterior allografts were prepared and fixed in appropriately sized tunnels created in 10 lb/ft 3 (0.16 g/cm3) dense synthetic bone blocks using screws of similar length and root and thread diameter designed with either a large buttress thread with a smaller taper or small buttress thread with a larger taper. Insertion torque was measured at one-third, two-thirds, and full screw insertion. After the graft fixation constructs were mounted in a servohydraulic-testing device with the loading axis aligned directly with the tunnel and preloaded to 25 N, they were cycled 3 times from 0 to 50 N, and then subjected to a 20 mm/minute traction force to failure.
RESULTS: All constructs failed by graft slippage past the screw. Mean maximum load at failure (360.5 +/- 68 N v 341.6 +/- 58 N, P = .2) and stiffness (63.6 +/- 16 N/mm 2 v 66.4 +/- 14 N/mm 2 , P = .89) was similar between constructs fixed with a large buttress thread small-taper screw and small buttress thread large-taper screw, respectively. The small buttress thread screw with a large taper displayed greater mean insertion torque at one-third insertion (4.1 +/- 0.57 in-lb v 3.2 +/- 0.49 in-lb, P = .03), whereas the large buttress thread screw with a small taper displayed greater mean insertion torque at full insertion (11.1 +/- 0.74 in-lb v 9.4 +/- 1.3 in-lb, P = .012). Mean differences were not observed at two-thirds screw insertion (P = .12).
CONCLUSIONS: Large buttress thread small-taper screws displayed biomechanical fixation characteristics comparable to small buttress thread large-taper screws. CLINICAL RELEVANCE: Given reports of superior screw-graft-bone tunnel contact area, these biomechanical results suggest that use of a large buttress screw with a small taper may be preferable for retrograde soft-tissue graft fixation.

Mesh:

Year:  2005        PMID: 16012494     DOI: 10.1016/j.arthro.2005.04.108

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  3 in total

1.  Does thread shape affect the fixation strength of the bioabsorbable interference screws for anterior cruciate ligament reconstructions? A biomechanical study.

Authors:  Gerardo L Garcés; Oscar Martel; Alejandro Yánez; Alberto Cuadrado
Journal:  BMC Musculoskelet Disord       Date:  2019-02-08       Impact factor: 2.362

2.  Osteolysis is observed around both bioabsorbable and nonabsorbable anchors on serial magnetic resonance images of patients undergoing arthroscopic rotator cuff repair.

Authors:  Ivan Micic; Erica Kholinne; Jae-Man Kwak; Kyoung-Hwan Koh; In-Ho Jeon
Journal:  Acta Orthop Traumatol Turc       Date:  2019-09-25       Impact factor: 1.511

3.  Redesigning Metal Interference Screws Can Improve Ease of Insertion While Maintaining Fixation of Soft-Tissue Anterior Cruciate Ligament Reconstruction Grafts.

Authors:  Kiron K Athwal; Breck R Lord; Piers E Milner; Alex Gutteridge; Andy Williams; Andrew A Amis
Journal:  Arthrosc Sports Med Rehabil       Date:  2020-04-06
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.