Literature DB >> 16006152

Validation of a motor activity system by a robotically controlled vehicle and using standard reference compounds.

John P Patterson1, Carrie G Markgraf, Maria Cirino, Alan S Bass.   

Abstract

INTRODUCTION: A series of experiments were undertaken to evaluate the accuracy, precision, specificity, and sensitivity of an automated, infrared photo beam-based open field motor activity system, the MotorMonitor v. 4.01, Hamilton-Kinder, LLC, for use in a good laboratory practices (GLP) Safety Pharmacology laboratory.
METHODS: This evaluation consisted of two phases: (1) system validation, employing known inputs using the EM-100 Controller Photo Beam Validation System, a robotically controlled vehicle representing a rodent and (2) biologic validation, employing groups of rats treated with the standard pharmacologic agents diazepam or D-amphetamine. The MotorMonitor's parameters that described the open-field activity of a subject were: basic movements, total distance, fine movements, x/y horizontal ambulations, rearing, and total rest time. These measurements were evaluated over a number of zones within each enclosure.
RESULTS: System validation with the EM-100 Controller Photo Beam Validation System showed that all the parameters accurately and precisely measured what they were intended to measure, with the exception of fine movements and x/y ambulations. Biologic validation using the central nervous system depressant diazepam at 1, 2, or 5 mg/kg, i.p. produced the expected dose-dependent reduction in rat motor activity. In contrast, the central nervous system stimulant D-amphetamine produced the expected increases in rat motor activity at 0.1 and 1 mg/kg, i.p, demonstrating the specificity and sensitivity of the system. DISCUSSION: Taken together, these studies of the accuracy, precision, specificity, and sensitivity show the importance of both system and biologic validation in the evaluation of an automated open field motor activity system for use in a GLP compliant laboratory.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16006152     DOI: 10.1016/j.vascn.2005.05.003

Source DB:  PubMed          Journal:  J Pharmacol Toxicol Methods        ISSN: 1056-8719            Impact factor:   1.950


  2 in total

1.  Delta opioid receptor ligands modulate anxiety-like behaviors in the rat.

Authors:  Shane A Perrine; Brian A Hoshaw; Ellen M Unterwald
Journal:  Br J Pharmacol       Date:  2006-04       Impact factor: 8.739

2.  Automated recording of home cage activity and temperature of individual rats housed in social groups: The Rodent Big Brother project.

Authors:  William S Redfern; Karen Tse; Claire Grant; Amy Keerie; David J Simpson; John C Pedersen; Victoria Rimmer; Lauren Leslie; Stephanie K Klein; Natasha A Karp; Rowland Sillito; Agis Chartsias; Tim Lukins; James Heward; Catherine Vickers; Kathryn Chapman; J Douglas Armstrong
Journal:  PLoS One       Date:  2017-09-06       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.