Hepatic arterial doppler sonography in patients with cirrhosis and controls: observer and equipment variability with use of the ultrasonic contrast agent SHU 508A.

G H Hübner1, N Steudel, A Reissmann, I Kühne, G Kleber, K Schärff, W E Fleig.   

Abstract

PURPOSE: The aims of this study on hepatic arterial Doppler sonography were to ascertain interobserver and interequipment variability, to investigate any potential artificial influence of the ultrasonic contrast agent on the Doppler measurements and to compare the results in healthy and cirrhotic subjects.
METHODS: Doppler sonography of the left hepatic artery was performed in nine healthy and nine cirrhotic subjects by three independent observers using three different devices. Continuous infusion of the ultrasonic contrast agent SHU 508A and placebo were administered in a double blind fashion. Systolic, mean and end diastolic peak velocities as well as resistive and pulsatility indices were measured.
RESULTS: Equipment associated variances (5.8 - 12.7 %) of the five Doppler parameters were greater than interobserver variances (0.3 - 3.6 %). No significant differences were observed between the velocities using ultrasonic contrast agent and placebo. Systolic (65.9 +/- 3.6 vs. 47.7 +/- 4.2 cm/s mean +/- SE, p = 0.02) and mean peak velocity (35.4 +/- 1.6 vs. 24.5 +/- 1.8 cm/s, p = 0.007) were significantly higher in cirrhotic than in healthy subjects whereas the resistive and pulsatility indices were not different.
CONCLUSIONS: Doppler sonography of the left hepatic artery performed by various observers is reproducible as long as the same device is used. Under clinical conditions, velocities are correctly measured with the use of ultrasonic contrast agent and are elevated in patients with cirrhosis.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16001345     DOI: 10.1055/s-2005-858103

Source DB:  PubMed          Journal:  Z Gastroenterol        ISSN: 0044-2771            Impact factor:   2.000


  2 in total

1.  Use of Caval Subtraction 2D Phase-Contrast MR Imaging to Measure Total Liver and Hepatic Arterial Blood Flow: Preclinical Validation and Initial Clinical Translation.

Authors:  Manil D Chouhan; Rajeshwar P Mookerjee; Alan Bainbridge; Simon Walker-Samuel; Nathan Davies; Steve Halligan; Mark F Lythgoe; Stuart A Taylor
Journal:  Radiology       Date:  2016-05-12       Impact factor: 11.105

2.  Influence of ultrasound contrast agents on spectral Doppler analysis in recipients of liver transplantation.

Authors:  Young Seo Cho; Kyoung Won Kim; Hye Young Jang; Bo Hyun Kim; Jeongjin Lee; Gi Won Song; Sung Gyu Lee; Dagvasumberel Munkhbaatar
Journal:  Clin Mol Hepatol       Date:  2017-07-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.