Literature DB >> 15987968

Hepatic hemorrhage caused by percutaneous tumor ablation: radiofrequency ablation versus cryoablation in a porcine model.

Sarah A Shock1, Paul F Laeseke, Lisa A Sampson, William D Lewis, Thomas C Winter, Jason P Fine, Fred T Lee.   

Abstract

PURPOSE: To determine the extent of hepatic hemorrhage caused by percutaneous cryoablation performed with a small-diameter cryoablation probe compared with that caused by percutaneous radiofrequency (RF) ablation in a porcine model.
MATERIALS AND METHODS: The study was pre-approved by the institutional research animal care and use committee, and husbandry and experiments complied with National Institutes of Health standards for care and use of laboratory animals. Percutaneous hepatic ablation was performed in 18 domestic pigs (mean weight, 45 kg) by using a 17-gauge (1.5-mm-diameter) RF electrode (n = 6), a cluster of three RF electrodes (n = 6), or a 13-gauge (2.4 mm-diameter) cryoprobe (n = 6). Ablation was performed in four sites per liver. Total blood loss, minimum lesion diameter, maximum lesion diameter, and lesion volume were determined for each group and compared by using analysis of variance.
RESULTS: Mean blood loss was 11.11 mL +/- 11.47 (standard deviation), 105.29 mL +/- 175.58, and 28.06 mL +/- 30.97 with the single RF electrode, RF electrode cluster, and cryoablation probe, respectively. Mean minimum and maximum lesion diameters were largest with the RF electrode cluster (2.40 and 3.98 cm, respectively), followed by the cryoablation probe (2.38 and 3.94 cm) and single RF electrode (1.49 and 2.63 cm). Mean minimum and maximum lesion diameters were significantly different between the single RF electrode and the RF electrode cluster, as well as between the single RF electrode and the cryoablation probe (P < .001). Mean lesion volume was largest for the RF electrode cluster (24.03 cm3), followed by those for the cryoablation probe (17.46 cm3) and single RF electrode (9.05 cm3) (single RF electrode vs cryoablation probe, P < .05). Lesion volumes were not significantly different with the RF electrode cluster versus the single RF electrode (P = .052) or with the RF electrode cluster versus the cryoablation probe (P = .381).
CONCLUSION: Mean blood loss from percutaneous cryoablation in this model was between that for RF ablation with the single electrode and that for RF ablation with the electrode cluster. Copyright RSNA, 2005

Entities:  

Mesh:

Year:  2005        PMID: 15987968     DOI: 10.1148/radiol.2361040533

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  6 in total

1.  Combined magnetic resonance imaging and optical surgical navigation system guidance of percutaneous liver cryoablation in a porcine model.

Authors:  Kang Zhou; Zhuoli Zhang; Matteo Figini; Junjie Shangguan; Yining Wang; Haifeng Shi; Yumei Li; Zhengyu Jin; Jie Pan
Journal:  Am J Transl Res       Date:  2018-05-15       Impact factor: 4.060

2.  Contrast-enhanced ultrasound (CEUS) follow-up after radiofrequency ablation or cryoablation of focal liver lesions: treated-area patterns and their changes over time.

Authors:  Aymeric Guibal; Caroline Bertin; Sophie Egels; Eric Savier; Philippe A Grenier; Olivier Lucidarme
Journal:  Eur Radiol       Date:  2012-11-09       Impact factor: 5.315

3.  Current tumor ablation technologies: basic science and device review.

Authors:  David F Saldanha; Vishal L Khiatani; Tami C Carrillo; Felix Y Yap; James T Bui; M Grace Knuttinen; Charles A Owens; Ron C Gaba
Journal:  Semin Intervent Radiol       Date:  2010-09       Impact factor: 1.513

Review 4.  Image-guided ablation of primary liver and renal tumours.

Authors:  David J Breen; Riccardo Lencioni
Journal:  Nat Rev Clin Oncol       Date:  2015-01-20       Impact factor: 66.675

5.  Cryotherapy is associated with improved clinical outcomes of sorafenib for the treatment of advanced hepatocellular carcinoma.

Authors:  Yongping Yang; Yinying Lu; Chunping Wang; Wenlin Bai; Jianhui Qu; Yan Chen; Xiujuan Chang; Linjing An; Lin Zhou; Zhen Zeng; Min Lou; Jiyun Lv
Journal:  Exp Ther Med       Date:  2011-12-01       Impact factor: 2.447

6.  Outcomes of ultrasound-guided percutaneous argon-helium cryoablation of hepatocellular carcinoma.

Authors:  Yongping Yang; Chunping Wang; Yinying Lu; Wenlin Bai; Linjing An; Jianhui Qu; Xudong Gao; Yan Chen; Lin Zhou; Yu Wu; Yongyi Feng; Minna Zhang; Xiujuan Chang; Jiyun Lv
Journal:  J Hepatobiliary Pancreat Sci       Date:  2012-11       Impact factor: 7.027

  6 in total