| Literature DB >> 15987130 |
Miguel Angel González-Martínez1, Eva María Brun, Rosa Puchades, Angel Maquieira, Kristy Ramsey, Fernando Rubio.
Abstract
A fully automated immunosensor for the herbicide glyphosate has been developed on the basis of the immunocomplex capture assay protocol. The sensor carries out on-line analyte derivatization prior to the assay and uses a selective anti-glyphosate serum, a glyphosate peroxidase enzyme tracer, and fluorescent detection. Under optimal conditions, the detection limit achieved is 0.021 microg/L with an analysis rate of 25 min per assay, autonomy of more than 48 h, and sensor reusability >500 analytical cycles. The immunosensor is able to discriminate structurally related molecules, such as aminomethylphosphonic acid, the main metabolite of glyphosate, and other related herbicides, such as glufosinate and glyphosine. Interferences from naturally occurring species (anions, cations, and humic substances) and their elimination were also studied. The immunosensor has been successfully applied to water and soil sample analysis, with good recoveries at levels lower than 1 microg/L. Results obtained with the immunosensor correlate well with data from a magnetic particle ELISA and LC/LC/MS chromatographic method.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15987130 DOI: 10.1021/ac048431d
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986