Scalp electrode placement by EC2 adhesive paste in long-term video-EEG monitoring.

Carolina Falco1, Fabio Sebastiano, Lucia Cacciola, Fabio Orabona, Raffaella Ponticelli, Pamela Stirpe, Giancarlo Di Gennaro.   

Abstract

OBJECTIVE: To evaluate the usefulness of an adhesive paste named EC2 (Grass-Telefactor) in comparison with collodion, for scalp electrodes placement in patients with drug resistant partial epilepsy monitored by long-term video-EEG.
METHODS: A total of 40 patients with drug resistant partial epilepsy participated in the study. In 20 patients, electrode placement on the scalp was made with collodion (group C) whereas in the remaining patients EC2 was used (Group P). After the electrode placement (T1) and after 24 h of recording (T2), the impedance of the electrodes was measured. Moreover, the time required to apply the electrodes and for their daily maintenance was calculated and recorded for all patients who entered the study.
RESULTS: At each observation, group C showed mean values of electrode impedance significantly higher that the group P (T1: 16.8 k omega; T2: 6.5 k omega vs T1: 2.4 k omega; T2: 4.0 k omega, respectively) (P < 1 x 10(-5)). The time required to make the montage and to provide its daily maintenance was significantly shorter in group P than in group C [20.8 and 10.5 min vs 44.3 and 19.7 min, respectively (P < 1 x 10(-5))].
CONCLUSIONS: We found that the use of EC2 paste in scalp electrode attachment is less time consuming, with better recording quality as a result of lower electrode impedance values, than the use of collodion. SIGNIFICANCE: EC2 paste may substitute collodion in electrode placement for long-term video-EEG monitoring, with an optimal cost-benefit ratio in terms of recording performance, time consumption, and safety.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15982926     DOI: 10.1016/j.clinph.2005.04.012

Source DB:  PubMed          Journal:  Clin Neurophysiol        ISSN: 1388-2457            Impact factor:   3.708


  6 in total

1.  The utility of conductive plastic electrodes in prolonged ICU EEG monitoring.

Authors:  Rohit R Das; Brendan P Lucey; Sherry H-Y Chou; Patricio S Espinosa; Amir A Zamani; Barbara A Dworetzky; Edward B Bromfield; Jong Woo Lee
Journal:  Neurocrit Care       Date:  2008-09-20       Impact factor: 3.210

2.  Characteristics of obstructive sleep apnea in myasthenia gravis patients: a single center study.

Authors:  Sung Jae Heo; Jin-Sun Jun; Donghwi Park; Ho-Won Lee; Jung-Soo Kim; Jin-Sung Park
Journal:  Neurol Sci       Date:  2019-01-16       Impact factor: 3.307

3.  A novel hydrogel electrolyte extender for rapid application of EEG sensors and extended recordings.

Authors:  Killian Kleffner-Canucci; Phan Luu; John Naleway; Don M Tucker
Journal:  J Neurosci Methods       Date:  2011-11-28       Impact factor: 2.390

4.  Single trial classification of motor imagination using 6 dry EEG electrodes.

Authors:  Florin Popescu; Siamac Fazli; Yakob Badower; Benjamin Blankertz; Klaus-R Müller
Journal:  PLoS One       Date:  2007-07-25       Impact factor: 3.240

5.  Unobtrusive ambulatory EEG using a smartphone and flexible printed electrodes around the ear.

Authors:  Stefan Debener; Reiner Emkes; Maarten De Vos; Martin Bleichner
Journal:  Sci Rep       Date:  2015-11-17       Impact factor: 4.379

6.  Comparison of artifacts between paste and collodion method of electrode application in pediatric EEG.

Authors:  Danielle Brigham; Yash Shah; Kanwaljit Singh; Ivan Pavkovic; Shefali Karkare; Sanjeev V Kothare
Journal:  Clin Neurophysiol Pract       Date:  2019-11-30
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.