The threat of adhesive embolization: BioGlue leaks through needle holes in aortic tissue and prosthetic grafts.

Scott A LeMaire1, Stacey A Carter, Taehee Won, Xinwen Wang, Lori D Conklin, Joseph S Coselli.   

Abstract

BACKGROUND: BioGlue Surgical Adhesive (CryoLife, Inc, Kennesaw, GA) is used to reinforce anastomoses during cardiovascular operations. Previous reports have raised concerns that adhesives may leak through suture-line needle holes and that resulting intraluminal glue may embolize. The purpose of this study was to determine if BioGlue leaks through anastomotic needle holes in aortic tissue and two common prosthetic graft materials.
METHODS: Polypropylene suture was used to create end-to-end anastomoses in gelatin-sealed woven polyester grafts (n = 45), expanded polytetrafluoroethylene (ePTFE) grafts (n = 45), and fresh porcine aortas (n = 45). An additional 45 anastomoses were created in ePTFE grafts using ePTFE sutures. The outer surface of each anastomosis was covered with BioGlue. Anastomoses underwent inspection with direct magnification or histology.
RESULTS: BioGlue leaked through needle holes and into the lumen in 10% of anastomoses (18 of 180). Leaks were significantly more common in fresh aorta (10 of 45, 22%) than in prosthetic grafts (8 of 135, 6%; p = 0.003). Suture size did not significantly affect the incidence of leak. The use of ePTFE sutures did not eliminate BioGlue leakage. Prosthetic graft leaks created discreet round adhesive particles. In contrast, aortic tissue leaks resulted in thin, friable flakes of glue extending along the intimal surface. Aortic histology confirmed that BioGlue reached the vessel lumen via the suture channels.
CONCLUSIONS: BioGlue leaked through the needle holes in fresh aortic tissue and prosthetic grafts. Intraluminal adhesive particles were easily dislodged, supporting concerns regarding embolization. The potential for adhesive embolization should be a factor when considering the relative risks and benefits of using BioGlue.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15975350     DOI: 10.1016/j.athoracsur.2005.02.004

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  9 in total

1.  BioGlue in 2011: what is its role in cardiac surgery?

Authors:  Castigliano M Bhamidipati; Joseph S Coselli; Scott A LeMaire
Journal:  J Extra Corpor Technol       Date:  2012-03

Review 2.  The use of surgical glue in acute type A aortic dissection.

Authors:  Shinichi Suzuki; Munetaka Masuda; Kiyotaka Imoto
Journal:  Gen Thorac Cardiovasc Surg       Date:  2013-11-21

3.  Mechanical characterization of a bifunctional Tetronic hydrogel adhesive for soft tissues.

Authors:  Lindsey Sanders; Roland Stone; Kenneth Webb; Thompson Mefford; Jiro Nagatomi
Journal:  J Biomed Mater Res A       Date:  2014-08-20       Impact factor: 4.396

4.  Cerebral embolic shower secondary to BioGlue use in triple valve surgery.

Authors:  Venkatesa Kumar Anakaputhur Rajan; Srikanth Kasturi; Niranjan Soundararajan; Priyankar Sinha
Journal:  Indian J Thorac Cardiovasc Surg       Date:  2022-03-30

5.  BioGlue cerebral embolism following acute type A aortic dissection repair.

Authors:  Manabu Yamasaki; Kohei Abe; Ryota Nakamura; Rihito Tamaki; Hiroyasu Misumi
Journal:  J Cardiol Cases       Date:  2022-07-16

6.  Results from a First-in-Human Trial of a Novel Vascular Sealant.

Authors:  Hans-Joachim Florek; Jan Brunkwall; Karl-Heinz Orend; Ian Handley; John Pribble; Ronald Dieck
Journal:  Front Surg       Date:  2015-07-01

7.  Bovine Serum Albumin-Glutaraldehyde Sealed Fish-Mouth Closure of the Pancreatic Remnant during Distal Pancreatectomy.

Authors:  Fritz Klein; Igor Maximilian Sauer; Johann Pratschke; Marcus Bahra
Journal:  HPB Surg       Date:  2017-01-17

8.  "How I do it: utilization of high-pressure sealants in aortic reconstruction".

Authors:  John A Elefteriades
Journal:  J Cardiothorac Surg       Date:  2009-06-26       Impact factor: 1.637

9.  Usefulness of a new gelatin glue sealant system for dural closure in a rat durotomy model.

Authors:  Hisashi Kawai; Ichiro Nakagawa; Fumihiko Nishimura; Yasushi Motoyama; Young-Su Park; Mitsutoshi Nakamura; Hiroyuki Nakase; Shuko Suzuki; Yoshito Ikada
Journal:  Neurol Med Chir (Tokyo)       Date:  2014-07-28       Impact factor: 1.742