Several radiation surveys, at medical linear accelerator facilities where lead or steel had been used with concrete to fabricate the primary barriers, revealed the existence of a sizable neutron field outside the shielding. This neutron field is produced by photoneutrons generated in the metal portion of the shield when the primary x-ray beam is aimed at the barrier. A method was developed to calculate the neutron dose-equivalent rate expected outside a primary shield when it is irradiated by a high-energy x-ray beam. It was found that the minimum photoneutron dose was produced when the metal part of the shield was positioned inside the treatment room in front of the concrete and also by using steel in place of lead. A thickness of less than or equal to 17 cm of metal on the inner surface of the shield produced only a slight increase in the neutron dose equivalent outside the barrier.Mesh:
Year: 1992 PMID: 1597389 DOI: 10.1097/00004032-199204000-00011
Source DB: PubMed Journal: Health Phys ISSN: 0017-9078 Impact factor: 1.316