Non-invasive assessment of coronary artery stent patency with multislice CT: preliminary experience.

F Cademartiri1, R Marano, G Runza, N Mollet, K Nieman, G Luccichenti, M Gualerzi, L Brambilla, P Coruzzi, M Galia, M Midiri.   

Abstract

PURPOSE: To evaluate the diagnostic accuracy of multislice computed tomography coronary angiography (MSCT-CA) in the detection of in-stent restenosis.
MATERIALS AND METHODS: Forty-two patients (33 male, 9 female, mean age 58+/-8 years) previously subjected to percutaneous implantation of coronary stent with suspected in-stent restenosis, underwent a 16-row MSCT (Sensation 16, Siemens) examination. The average time between stent implantation and MSCT-CA was 7.4+/-5.3 months. The following scan parameters were used: collimation 16x0.75 mm, rotation time 0.42 s, feed 3.0 mm/rot., kV 120, mAs 500. After administration of iodinated contrast material (Iomeprol 400 mgI/ml, 100 ml at 4 ml/s) and bolus chaser (40 ml of saline at 4 ml/s) the scan was completed in <20 s. All segments with a stent were assessed by two observers in consensus and were graded according to the following scheme: patent stent, in-stent intimal hyperplasia (IIH) (lumen reduction <50%), in-stent restenosis (ISR) (=/>50%), in-stent occlusion (ISO) (100%). Consensus reading was compared with coronary angiography.
RESULTS: Forty-seven stents were assessed (16 in the right coronary artery; 4 in the left main; 22 in the left anterior descending; 5 in the circumflex). In 7 (17%) stents there was ISR (3) or ISO (4), and in 4 (10%) stents there was IIH. The sensitivity and negative predictive values for the detection of ISO were 80% and 98%, respectively, while for the detection of ISR+ISO they were 50% and 89%, respectively.
CONCLUSIONS: Although the results are encouraging, the follow-up of stent patency with MSCT-CA does not show a diagnostic accuracy suitable for clinical implementation.

Entities:  

Year:  2005        PMID: 15973223

Source DB:  PubMed          Journal:  Radiol Med        ISSN: 0033-8362            Impact factor:   3.469


  8 in total

1.  Evaluation of in-stent restenosis in proximal coronary arteries with multidetector computed tomography (MDCT).

Authors:  V Chabbert; D Carrie; M Bennaceur; E Maupas; V Lauwers; M Mhem; T Lhermusier; M Elbaz; F Joffre; H Rousseau; J Puel
Journal:  Eur Radiol       Date:  2006-11-18       Impact factor: 5.315

Review 2.  Cardiac CT: coronary arteries and beyond.

Authors:  Andreas H Mahnken; Georg Mühlenbruch; Rolf W Günther; Joachim E Wildberger
Journal:  Eur Radiol       Date:  2006-10-26       Impact factor: 5.315

3.  In vitro evaluation of coronary stents and in-stent stenosis using a dynamic cardiac phantom and a 64-detector row CT scanner.

Authors:  T Schlosser; T Scheuermann; S Ulzheimer; O K Mohrs; M Kühling; P E Albrecht; T Voigtländer; J Barkhausen; A Schmermund
Journal:  Clin Res Cardiol       Date:  2007-08-21       Impact factor: 5.460

4.  Comparison of 16-slice computed tomography with conventional angiography to evaluate coronary artery stent patency.

Authors:  Irmak Durur-Subasi; Mecit Kantarci; Afak Durur-Karakaya; Adnan Okur
Journal:  Eurasian J Med       Date:  2009-04

5.  Diagnostic accuracy of 16-slice multidetector-row CT for detection of in-stent restenosis vs detection of stenosis in nonstented coronary arteries.

Authors:  Joelle M Kefer; Emmanuel Coche; Jean-Louis J Vanoverschelde; Bernhard L Gerber
Journal:  Eur Radiol       Date:  2006-05-30       Impact factor: 5.315

6.  Comparison between different kernel reformatting filters in 3D quantitative analysis of MDCT coronary angiography.

Authors:  R Malagò; A Pezzato; C Barbiani; W Mantovani; G Caliari; U Alfonsi; D Tavella; R Pozzi Mucelli
Journal:  Radiol Med       Date:  2011-09-02       Impact factor: 3.469

7.  Diagnostic accuracy of in-stent coronary restenosis detection with multislice spiral computed tomography: a meta-analysis.

Authors:  Michèle Hamon; Laure Champ-Rigot; Rémy Morello; John W Riddell; Martial Hamon
Journal:  Eur Radiol       Date:  2007-09-01       Impact factor: 5.315

8.  Diagnostic accuracy of 64-slice CT in the assessment of coronary stents.

Authors:  F Cademartiri; A Palumbo; E Maffei; L La Grutta; G Runza; F Pugliese; M Midiri; N R A Mollet; W B Meijboom; A Menozzi; L Vignali; C Reverberi; D Ardissino; G P Krestin
Journal:  Radiol Med       Date:  2007-06-11       Impact factor: 6.313

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.