| Literature DB >> 15967303 |
José Hernández1, G Marcelín-Jiménez, L Rivera, P Angeles Alionka, L Contreras, M Hinojosa, L Martínez-Rossier, O Amancio, A Fernández.
Abstract
Diphenidol was determined by an HPLC method developed in our laboratory. It was validated and proved to be linear in the 40-400 ng/ml range. Accuracy for quality-control samples for intra and inter day assays ranged from 96.1-98.9% and 98.8-101.4%, respectively. This method was applied to a multi-dose bioequivalence study. No serious side effects were observed in the multi-dose design. Pharmacokinetic parameters (mean+/-standard error [S.E.]) of Cavg (ng/ml) and AUCtau (ngh/ml) for reference and test products were 139.54+/-12.66 versus 148.60+/-16.51 and 551.07+/-53.53 versus 588.78+/-69.02, respectively. Log-transformed values were compared by analysis of variance (ANOVA) followed by the classical 90% confidence interval (CI 90%) test and Schuirmann's test. Confidence limits ranged from 91.48-116.18% for Cmax and from 91.24-117.65% for AUCtau. These results suggest that the analytical method was linear, precise, and accurate for our purpose, and that both assayed formulations were bioequivalent.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15967303 DOI: 10.1016/j.jpba.2005.01.037
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935