| Literature DB >> 15957599 |
James Revell1, Majid Mirmehdi, Donal McNally.
Abstract
We present the development and validation of an image based speckle tracking methodology, for determining temporal two-dimensional (2-D) axial and lateral displacement and strain fields from ultrasound video streams. We refine a multiple scale region matching approach incorporating novel solutions to known speckle tracking problems. Key contributions include automatic similarity measure selection to adapt to varying speckle density, quantifying trajectory fields, and spatiotemporal elastograms. Results are validated using tissue mimicking phantoms and in vitro data, before applying them to in vivo musculoskeletal ultrasound sequences. The method presented has the potential to improve clinical knowledge of tendon pathology from carpel tunnel syndrome, inflammation from implants, sport injuries, and many others.Entities:
Mesh:
Year: 2005 PMID: 15957599 DOI: 10.1109/TMI.2005.848331
Source DB: PubMed Journal: IEEE Trans Med Imaging ISSN: 0278-0062 Impact factor: 10.048