Literature DB >> 15955187

Electromagnetic interference of implantable unipolar cardiac pacemakers by an induction oven.

Minoru Hirose1, Mizuho Hida, Eiji Sato, Kenichi Kokubo, Masaki Nie, Hirosuke Kobayashi.   

Abstract

Induction ovens have been reported to exert electromagnetic interference on implanted cardiac pacemakers. In an attempt to quantitatively investigate the electromagnetic interference caused by an induction oven on implantable unipolar cardiac pacemakers, we measured the distribution profile of the magnetic field intensity, both with and without a pan on the induction oven. We also performed the inhibition test and asynchronous test using four kinds of pacemakers housed in the standardized Irnich human body model, and measured the maximal distance from the induction oven up to which the interference occurred. In the pan-detection mode of the oven in the absence of a pan, the distribution profile of the magnetic field intensity peaked at the center of the cooking plate, and during induction heating of a pan placed on the induction oven, it was the largest at the circular top-edge of the pan. Pacemaker pulses were inhibited by the induction oven, or generated by the reversion mechanism. The maximal interference distance from the oven was 34 cm for one of the pacemakers. Thus, the safe distance from an induction oven of a patient with an implanted cardiac pacemaker is considered to be 50 cm or more. In conclusion, in the pan-detection mode of the oven in the absence of a pan, the distribution profile of the magnetic field intensity peaked at the center of the cooking plate, and during the induction heating of a pan placed on the oven, it peaked at the circular edge of the pan. The induction oven asynchronized or generated pulses in implantable unipolar cardiac pacemakers up to a maximal distance of 34 cm from the induction oven.

Entities:  

Mesh:

Year:  2005        PMID: 15955187     DOI: 10.1111/j.1540-8159.2005.09565.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  5 in total

1.  Experimental study on malfunction of pacemakers due to exposure to different external magnetic fields.

Authors:  Maria Tiikkaja; Tommi Alanko; Harri Lindholm; Maila Hietanen; Juha Hartikainen; Lauri Toivonen
Journal:  J Interv Card Electrophysiol       Date:  2012-01-11       Impact factor: 1.900

2.  Electromagnetic interference of implantable cardiac devices from a shoulder massage machine.

Authors:  Saeko Yoshida; Kousaku Fujiwara; Satoshi Kohira; Minoru Hirose
Journal:  J Artif Organs       Date:  2014-04-08       Impact factor: 1.731

Review 3.  Mechanized scaling with ultrasonics: Perils and proactive measures.

Authors:  Rashmi Paramashivaiah; M L V Prabhuji
Journal:  J Indian Soc Periodontol       Date:  2013-07

4.  Testing of common electromagnetic environments for risk of interference with cardiac pacemaker function.

Authors:  Maria Tiikkaja; Aapo L Aro; Tommi Alanko; Harri Lindholm; Heli Sistonen; Juha E K Hartikainen; Lauri Toivonen; Jukka Juutilainen; Maila Hietanen
Journal:  Saf Health Work       Date:  2013-07-20

5.  Electromagnetic interference in cardiac electronic implants caused by novel electrical appliances emitting electromagnetic fields in the intermediate frequency range: a systematic review.

Authors:  Sarah Driessen; Andreas Napp; Kristina Schmiedchen; Thomas Kraus; Dominik Stunder
Journal:  Europace       Date:  2019-02-01       Impact factor: 5.214

  5 in total