Abstract

Metals, alloys, and poisoned food were analyzed with a hand-held X-ray fluorescence (XRF) spectrometer, with a shield (wrapping or casing material) inserted between these objects and the spectrometer, in order to examine the possibility of analyzing the contents of packages. Elements such as Fe, Cr, Ni, Cu, Zn, Pb, Mo, and As were detected in the objects. The fluorescent intensity of each element in the object decreased exponentially as the thickness of the shield increased, and the degree of decrease depended on both the material of the shield and the energy of fluorescent X-rays. The thickness of the shield can be calculated by using the intensity ratio Fe Kbeta/Kalpha or Pb Lbeta/Lalpha when the object is iron or lead, or by using the intensity of the Compton scattering of incident X-rays. The original peak intensity, i.e. intensity without a shield, of an element in an object can be estimated with the thickness of the shield obtained. Because the original peak intensity is calculated using an exponential function of the thickness of the shield, calculation of the intensity ratio, e.g. Zn Kalpha/Cu Kalpha for brass, is effective for cancelling the estimation error for the thickness of the shield. The composition of brass and steel can be estimated with an error of less than 30% by using the intensity of the Compton scattering.

Entities:  

Year:  2005        PMID: 15939161     DOI: 10.1016/j.forsciint.2005.02.017

Source DB:  PubMed          Journal:  Forensic Sci Int        ISSN: 0379-0738            Impact factor:   2.395


  3 in total

Review 1.  Substandard/counterfeit antimicrobial drugs.

Authors:  Theodoros Kelesidis; Matthew E Falagas
Journal:  Clin Microbiol Rev       Date:  2015-04       Impact factor: 26.132

2.  Evaluation of a New Optic-Enabled Portable XRF Instrument for Measuring Toxic Metals/Metalloids in Consumer Goods and Cultural Products.

Authors:  Diana Guimarães; Meredith L Praamsma; Patrick J Parsons
Journal:  Spectrochim Acta Part B At Spectrosc       Date:  2016-03-22       Impact factor: 3.752

3.  Paper analytical devices for fast field screening of beta lactam antibiotics and antituberculosis pharmaceuticals.

Authors:  Abigail A Weaver; Hannah Reiser; Toni Barstis; Michael Benvenuti; Debarati Ghosh; Michael Hunckler; Brittney Joy; Leah Koenig; Kellie Raddell; Marya Lieberman
Journal:  Anal Chem       Date:  2013-06-18       Impact factor: 6.986

  3 in total