Literature DB >> 15936604

Embolic protection devices for carotid artery stenting: is there a difference between filter and distal occlusive devices?

Ralf Zahn1, Thomas Ischinger, Bernd Mark, Sabine Gass, Uwe Zeymer, Wolfgang Schmalz, Klaus Haerten, Karl Eugen Hauptmann, Enz-Rüdiger von Leitner, Wolfgang Kasper, Ulrich Tebbe, Jochen Senges.   

Abstract

OBJECTIVES: We sought to compare the efficacy of a filter embolic protection device (F-EPD) and a distal occlusive embolic protection device (DO-EPD) in patients undergoing carotid artery stenting (CAS).
BACKGROUND: The embolic protection device (EPD) may lower the periprocedural rate of cerebral ischemic events during CAS. However, it is unclear whether there is a difference in effectiveness between the different types of EPD.
METHODS: We analyzed data from the Carotid Artery Stent (CAS) Registry.
RESULTS: From July 1996 to July 2003, 1,734 patients were included in the prospective CAS Registry. Of these patients, 729 patients were treated with an EPD, 553 (75.9%) with F-EPD, and 176 (24.1%) with DO-EPD. Patients treated with DO-EPD were more likely to be treated for symptomatic stenosis (64.5% vs. 53.4%, p = 0.011). The carotid lesions in patients treated under DO-EPD seemed to be more complicated, as expressed by a higher proportion of ulcers (p = 0.035), severe calcification (p = 0.039), a longer lesion length (p = 0.025), and a higher pre-interventional grade of stenosis (p < 0.001). The median duration of the CAS intervention was 30 min in the DO-EPD group, compared with 48 min in the filter group (p < 0.001). No differences in clinical events rate between the two groups of protection devices were observed. Multivariate analysis on the occurrence of the combined end point of in-hospital death or stroke found no difference between filter- and DO-EPD (4 of 176 [2.3%] for DO-EPD vs. 10 of 551 [1.8%] for F-EPD; adjusted odds ratio = 1.04, 95% confidence interval 0.24 to 4.44; p = 0.958).
CONCLUSIONS: Filter EPD is the currently preferred method of EPD in clinical practice. Both F-EPD and DO-EPD seem to be equally effective during CAS.

Entities:  

Mesh:

Year:  2005        PMID: 15936604     DOI: 10.1016/j.jacc.2005.02.067

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  11 in total

1.  [GeCAS registry. Sense and purpose of a carotid percutaneous transluminal angioplasty register in the era of obligatory quality assurance].

Authors:  N Werner; R Zahn
Journal:  Herz       Date:  2013-11       Impact factor: 1.443

2.  New ischemic brain lesions on diffusion-weighted MRI after carotid artery stenting with filter protection: frequency and relationship with plaque morphology.

Authors:  D Stojanov; M Ilic; P Bosnjakovic; M Zivkovic; S Jolic; N Vukasinovic; A Ignjatovic; B Ilic; D Benedeto-Stojanov
Journal:  AJNR Am J Neuroradiol       Date:  2011-12-22       Impact factor: 3.825

3.  Diffusion-weighted MR imaging in carotid angioplasty and stenting with protection by the reversed carotid arterial flow.

Authors:  F Asakura; K Kawaguchi; H Sakaida; N Toma; S Matsushima; K Kuraishi; H Tanemura; Y Miura; M Maeda; W Taki
Journal:  AJNR Am J Neuroradiol       Date:  2006-04       Impact factor: 3.825

4.  Use of embolic protective devices in treating acute arterial occlusions: an interventional radiology and vascular surgery collaborative learning experience.

Authors:  Joel Woodley-Cook; Vikram Prabhudesai; Tony Moloney
Journal:  BMJ Case Rep       Date:  2013-04-10

5.  Spasm induced by protection balloon during carotid artery stenting.

Authors:  Masanori Tsutsumi; Kiyoshi Kazekawa; Masanari Onizuka; Hiroshi Aikawa; Kouhei Nii; Tomonobu Kodama; Minoru Iko; Makoto Tomokiyo; Shuko Matsubara; Akira Tanaka
Journal:  Radiat Med       Date:  2007-08-27

6.  Distal protection filter device efficacy with carotid artery stenting: comparison between a distal protection filter and a distal protection balloon.

Authors:  Minoru Iko; Masanori Tsutsumi; Hiroshi Aikawa; Yoshihisa Matsumoto; Yoshinori Go; Kouhei Nii; Gorou Abe; Iwae Ye; Yasuyuki Nomoto; Kiyoshi Kazekawa
Journal:  Jpn J Radiol       Date:  2012-11-20       Impact factor: 2.374

7.  Double embolic protection during carotid artery stenting with persistent hypoglossal artery.

Authors:  Christine F Silva; Samuel Y Hou; Anna L Kühn; Richard H Whitten; Ajay K Wakhloo
Journal:  BMJ Case Rep       Date:  2013-05-02

8.  Cerebral ischemia detected with diffusion-weighted MR imaging after protected carotid artery stenting: comparison of distal balloon and filter device.

Authors:  Suk Jung Kim; Hong Gee Roh; Pyoung Jeon; Keon Ha Kim; Kwang Ho Lee; Hong Sik Byun; Won Jin Moon; Gyeong Moon Kim; Young Wook Kim; Dong Ik Kim
Journal:  Korean J Radiol       Date:  2007 Jul-Aug       Impact factor: 3.500

9.  Treatment outcomes of carotid artery stenting with two types of distal protection filter device.

Authors:  Minoru Iko; Hiroshi Aikawa; Yoshinori Go; Kanji Nakai; Masanori Tsutsumi; Iwae Yu; Taichiro Mizokami; Kimiya Sakamoto; Ritsuro Inoue; Takafumi Mitsutake; Ayumu Eto; Hayatsura Hanada; Kiyoshi Kazekawa
Journal:  Springerplus       Date:  2014-03-08

Review 10.  Carotid artery stenting: current and emerging options.

Authors:  Simon Morr; Ning Lin; Adnan H Siddiqui
Journal:  Med Devices (Auckl)       Date:  2014-10-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.