Determination of optimal gadolinium concentration using SSFP for catheter-directed contrast-enhanced coronary MR angiography.

Brian E Schirf1, Jordin D Green, Kent Sato, Andrew C Larson, Ali Shaibani, Riad Salem, Debiao Li, Reed A Omary.   

Abstract

RATIONALE AND
OBJECTIVES: To determine the optimal gadolinium concentration for catheter-directed coronary magnetic resonance angiography (MRA) using magnetization-prepared steady-state free-precession (SSFP) in swine.
MATERIALS AND METHODS: In six pigs, we performed real-time MR imaging-guided coronary artery catheterization using a 1.5 T MR scanner. For catheter-directed coronary MRA, we injected 3-4 mL of dilute Gd at 1 mL/second for each tested concentration (4%, 8%, 10%, and 12% Gd). Eleven images per concentration were acquired using electrocardiographic-triggered, magnetization-prepared two-dimensional (2D) projection SSFP. We compared mean relative signal-to-noise ratio (SNR) values for each concentration using two-way analysis of variance.
RESULTS: The targeted coronary artery was catheterized under real-time MR guidance in all pigs. Magnetization-prepared 2D projection SSFP successfully depicted the coronary arteries in all 44 injections. Mean relative SNR (+/- standard error) was 7.2 +/- 0.49 for 4%, 8.8 +/- 0.47 for 8%, 9.5 +/- 0.38 for 10%, and 8.8 +/- 0.41 for 12%. Injections of 4% dilute gadolinium yielded significantly less relative SNR than the other tested concentrations (P < .05). There were no statistically significant differences between the remaining concentrations.
CONCLUSION: For catheter-directed contrast-enhanced coronary MRA, the ideal gadolinium concentration should maximize relative SNR and limit the total gadolinium dose. Using these criteria, of those concentrations we tested in the swine model, 8% injected gadolinium was superior for catheter-directed SSFP coronary MRA.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15935975     DOI: 10.1016/j.acra.2005.02.013

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  3 in total

1.  Comparison of X-ray fluoroscopy and interventional magnetic resonance imaging for the assessment of coronary artery stenoses in swine.

Authors:  Jordin D Green; Reed A Omary; Brian E Schirf; Richard Tang; Biao Lu; James A Gehl; J Jenny Huang; James C Carr; F Scott Pereles; Debiao Li
Journal:  Magn Reson Med       Date:  2005-11       Impact factor: 4.668

2.  Three-dimensional contrast-enhanced steady-state free precession for improved catheter-directed coronary magnetic resonance angiography.

Authors:  Jordin D Green; Reed A Omary; Brian E Schirf; Richard Tang; James C Carr; Debiao Li
Journal:  J Magn Reson Imaging       Date:  2005-09       Impact factor: 4.813

3.  Accuracy of contrast-enhanced cine-MR sequences in the assessment of left ventricular function: comparison with precontrast cine-MR sequences. Results of a bicentric study.

Authors:  Jean-Christophe Lasalarie; Jean-Michel Serfaty; Christophe Carre; David Messika-Zeitoun; Christophe Jeannot; Elisabeth Schouman-Claeys; Jean-Pierre Laissy
Journal:  Eur Radiol       Date:  2007-05-08       Impact factor: 5.315

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.