Threshold pixel size for shape determination of microcalcifications in digital mammography: a pilot study.

M Ruschin1, B Hemdal, I Andersson, S Börjesson, M Håkansson, M Båth, A Grahn, A Tingberg.   

Abstract

The effect of pixel size on shape determination in screening digital mammography systems was studied using a shape identification task as the measured outcome. Ten microcalcifications on screen-films were digitised to a range of pixel sizes (2.5-200 microm) and extracted from computed radiography (CR) images (50 microm) acquired under equivalent imaging conditions. Fifteen observers attempted to identify the shape of each microcalcification at each pixel size. The results were collated to provide a fraction of correct responses vs. pixel size curve for each microcalcification. Averaging over all shapes, pixel values >100 microm lead to a significant decrease in shape determination ability (p < 0.01) for digitised screen-film. For CR images, half the shapes were not properly identified. Hence, although 20-100 microm was sufficient for microcalcification shape determination for digitised screen-film images, 50 microm was only borderline sufficient for the CR digital images.

Entities:  

Mesh:

Year:  2005        PMID: 15933149     DOI: 10.1093/rpd/nch561

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  3 in total

1.  Detection of simulated microcalcifications in a phantom with digital mammography: effect of pixel size.

Authors:  Sankararaman Suryanarayanan; Andrew Karellas; Srinivasan Vedantham; Ioannis Sechopoulos; Carl J D'Orsi
Journal:  Radiology       Date:  2007-05-23       Impact factor: 11.105

Review 2.  Digital mammography: what do we and what don't we know?

Authors:  Ulrich Bick; Felix Diekmann
Journal:  Eur Radiol       Date:  2007-02-14       Impact factor: 5.315

3.  Should previous mammograms be digitised in the transition to digital mammography?

Authors:  S Taylor-Phillips; M G Wallis; A G Gale
Journal:  Eur Radiol       Date:  2009-03-18       Impact factor: 5.315

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.