A Grahn1, B Hemdal, I Andersson, M Ruschin, A Thilander-Klang, S Börjesson, A Tingberg, S Mattsson, M Håkansson, M Båth, L G Månsson, J Medin, F Wanninger, W Panzer.
Abstract
The European Commission (EC) quality criteria for screen-film mammography are used as a tool to assess image quality. A new set of criteria was developed and initially tested in a previous study. In the present study, these criteria are further evaluated using screen-film mammograms that have been digitised, manipulated to simulate different image quality levels and reprinted on film. Expert radiologists have evaluated these manipulated images using both the original (EC) and the new criteria. A comparison of three different simulated dose levels reveals that the new criteria yield a larger separation of image criteria scores than the old ones. These results indicate that the new set of image quality criteria has a higher discriminative power than the old set and thus seems to be more suitable for evaluation of image quality in mammography.Mesh:
Year: 2005 PMID: 15933143 DOI: 10.1093/rpd/nch573
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972