Scatter, spatial resolution, and quantitative recovery in high resolution SPECT.

Z Szabo1, J M Links, C Seki, J Rhine, H N Wagner.   

Abstract

The potential use of single photon emission CT (SPECT) for quantification depends on its physical performance characteristics. We investigated the performance of a high resolution four-head brain SPECT scanner (Neuro-Spect; Summit/Hitachi). With an attenuation coefficient of 0.11 cm-1 and the Chang correction method, the calibration factor of the scanner was 515 (cpm/ml)/(microCi/ml) and showed only a minimal but systematic dependence on object size. Without scatter, the resolution was 4.7 mm (full width at half-maximum); in a scatter medium, the resolution was 5.3-10.0 mm with high resolution collimation and 7.7-18.8 mm with general purpose collimation, depending on filtering. A recovery coefficient of close to 100% was measured in the center of spheres with a size of at least 20 mm placed in a cylindrical water-filled phantom. In lesions of this size, scatter was 20%. We conclude from our measurements that the investigated high resolution SPECT offers significant improvements in resolution, scatter, and recovery, which will improve both the quality of brain studies and the measurement of quantitative parameters such as the cortex/white matter ratio.

Entities:  

Mesh:

Year:  1992        PMID: 1592932     DOI: 10.1097/00004728-199205000-00022

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  2 in total

Review 1.  Understanding Fourier space and filter selection.

Authors:  I G Zubal; G Wisniewski
Journal:  J Nucl Cardiol       Date:  1997 May-Jun       Impact factor: 5.952

Review 2.  Scatter correction in scintigraphy: the state of the art.

Authors:  I Buvat; H Benali; A Todd-Pokropek; R Di Paola
Journal:  Eur J Nucl Med       Date:  1994-07
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.