| Literature DB >> 15925218 |
Saly Romero-Torres1, José D Pérez-Ramos, Kenneth R Morris, Edward R Grant.
Abstract
We report new results suggesting the feasibility of Raman spectrometry as a tool by which to examine the variability of tablet coatings. Our experiments feature a probe that can operate with a revolving laser focus to average content and coating non-uniformity. Raman spectral changes are correlated with tablet exposure times in a pan coater by means of partial least squares (PLS) multivariate analysis. Statistical models are found to be improved by pre-processing schemes that emphasize spectral changes while minimizing the effects of background light scattering and fluorescence. These pre-processing techniques include multiplicative scatter correction (MSC) and standard normal variate (SNV) transformation, used in concert with Savitzky-Golay second derivative smoothing (SGSD). The two approaches give comparable results yielding R2 values for PLS calibration and cross-calibrated prediction variance regression of 0.999 and 0.997, respectively. Correlation results and model residual values demonstrate that Raman spectroscopy serves sensitively to reflect the coating thickness of the tablets studied.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15925218 DOI: 10.1016/j.jpba.2005.01.007