Stability of biodegradable implants in treatment of mandibular fractures.

Kaan C Yerit1, Sibylle Hainich, Dritan Turhani, Clemens Klug, Gert Wittwer, Michael Ockher, Oliver Ploder, Gerhard Undt, Arnulf Baumann, Rolf Ewers.   

Abstract

BACKGROUND: Biodegradable implants have not been used on a large scale for internal fixation of mandibular fractures because of presumed inferior mechanical properties. This prospective clinical trial was designed to elucidate the stability and biocompatibility of self-reinforced poly-L/D-lactide plates and screws used to stabilize a variety of mandible fractures by open reduction and internal fixation.
METHODS: Sixty-six consecutive patients (22 female, 44 male; mean age, 23.9 years) with a total of 89 fractures at various sites of the mandible were included in the study. Stability of plates and screws and bone healing were observed by clinical and radiographic assessment. Intermaxillary fixation was applied in eight patients with concomitant condylar fractures for 2 to 3 weeks.
RESULTS: The self-reinforcement technique provided sufficient mechanical stability of the implants for primary healing of these high-load mandibular bone areas. Postoperative complications were transient and limited to wound dehiscence and localized wound infection (two patients). In some patients, hypesthesia (three patients) or slight pain (10 patients) was reported at the 1-year recall examination, but implant-related serious adverse tissue reactions were not observed during the follow-up (mean, 24.4 months; range 6.4 to 44.3 months).
CONCLUSIONS: On the basis of these preliminary results, the authors conclude that biodegradable self-reinforced implants show efficient stability during initial bone healing and promise a high potential for successful use in osteofixation of mandibular fractures.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15923830     DOI: 10.1097/01.prs.0000165075.51898.6f

Source DB:  PubMed          Journal:  Plast Reconstr Surg        ISSN: 0032-1052            Impact factor:   4.730


  9 in total

1.  Choice of internal rigid fixation materials in the treatment of facial fractures.

Authors:  Mirko S Gilardino; Elliot Chen; Scott P Bartlett
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2009-03

2.  Extrusion forces of resorbable tacks and titanium screws in laryngeal chondrosynthesis.

Authors:  Norbert H Kleinsasser; Peter J Wild; J Carmelo Pérez Alvarez; Till O Seidler; Christian Schurr; Katrin Froelich; Tamás Hacki; Ferdinand Hofstaedter; Matthias A Graf; Rudolf Hagen
Journal:  Eur Arch Otorhinolaryngol       Date:  2006-11-23       Impact factor: 2.503

3.  Resorbable Versus Titanium Rigid Fixation for Pediatric Mandibular Fractures: A Systematic Review, Institutional Experience and Comparative Analysis.

Authors:  Matthew E Pontell; Eva B Niklinska; Stephane A Braun; Nolan Jaeger; Kevin J Kelly; Michael S Golinko
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2021-06-21

4.  Clinical and radiographic evaluation of biodegradable bone plates in the treatment of mandibular body fractures.

Authors:  Sherin Kamal Elhalawany; Bassel Tarakji; Sn Azzeghaiby; Ibrahim Alzoghaibi; Kusai Baroudi; Mohammad Zakaria Nassani
Journal:  Niger Med J       Date:  2015 Jan-Feb

Review 5.  Hardware Removal in Craniomaxillofacial Trauma: A Systematic Review of the Literature and Management Algorithm.

Authors:  Thomas J Cahill; Rikesh Gandhi; Alexander C Allori; Jeffrey R Marcus; David Powers; Detlev Erdmann; Scott T Hollenbeck; Howard Levinson
Journal:  Ann Plast Surg       Date:  2015-11       Impact factor: 1.539

6.  In Silico Biomechanical Evaluation of WE43 Magnesium Plates for Mandibular Fracture Fixation.

Authors:  Vincenzo Orassi; Heilwig Fischer; Georg N Duda; Max Heiland; Sara Checa; Carsten Rendenbach
Journal:  Front Bioeng Biotechnol       Date:  2022-02-10

Review 7.  Titanium or Biodegradable Osteosynthesis in Maxillofacial Surgery? In Vitro and In Vivo Performances.

Authors:  Barzi Gareb; Nico B Van Bakelen; Arjan Vissink; Ruud R M Bos; Baucke Van Minnen
Journal:  Polymers (Basel)       Date:  2022-07-07       Impact factor: 4.967

8.  Clinical courses and degradation patterns of absorbable plates in facial bone fracture patients.

Authors:  Young Min Kim; Jong Hun Lee
Journal:  Arch Craniofac Surg       Date:  2019-10-20

9.  Comparison of the mechanical properties of biodegradable and titanium osteosynthesis systems used in oral and maxillofacial surgery.

Authors:  Barzi Gareb; Charlotte C Roossien; Nico B van Bakelen; Gijsbertus J Verkerke; Arjan Vissink; Ruud R M Bos; Baucke van Minnen
Journal:  Sci Rep       Date:  2020-10-23       Impact factor: 4.379

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.