Characterization of ordered-subsets expectation maximization with 3D post-reconstruction Gauss filtering and comparison with filtered backprojection in 99mTc SPECT.

Marco Brambilla1, Barbara Cannillo, Marco Dominietto, Lucia Leva, Chiara Secco, Eugenio Inglese.   

Abstract

PURPOSE: To characterize ordered-subset expectation maximization algorithm with a fixed 3D Gauss post-reconstruction filtering (OSEM) in 99mTc SPECT as for noise, contrast and spatial resolution with varying number of subset and iteration and to compare OSEM with an optimized set of parameters, with filtered backprojection (FBP) with filter parameters typical of brain and myocardial SPECT, both with and without Chang's method of attenuation correction (AC).
METHODS: SPECT images of a Jaszczak phantom with cold rod inserts, hot and cold spheres and capillary line sources were acquired. Different background activity concentrations of the phantom were simulated as well as different lesion-to-background activity ratios. OSEM reconstructions were halted after 5, 10 and 15 iterations using 4, 8 and 16 subsets.
RESULTS: The effect of subset and iteration number over noise is additive: thus, it is possible to define an EM-equivalent iteration number that indicates the product between the subset and the iteration numbers. Noise increases linearly with increasing EM-equivalent iteration number. For each level of nominal contrast, the measured contrast after OSEM shows a little increase with increasing iteration number and saturates after 80 EM-equivalent iterations. The application of AC leads to diminished contrast values both in FBP and OSEM. The contrast of cold lesions after OSEM increases with increasing number of EM-equivalent iteration number: after 80 iterations the contrast values with OSEM overtake the ones obtained with FBP; contrast values diminished as background concentration raised. Resolution values did not change with increasing EM-equivalent iteration number and were higher than those obtained with FBP.
CONCLUSION: The major findings of the present work are the demonstration of additivity of subset and iteration in OSEM over noise, with the possibility of defining an EM equivalent iteration number, and the superiority of OSEM with respect to FBP in terms of spatial resolution.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15909485     DOI: 10.1007/BF03027384

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  8 in total

1.  Feasibility and operator variability of myocardial blood flow and reserve measurements with ⁹⁹mTc-sestamibi quantitative dynamic SPECT/CT imaging.

Authors:  Ran Klein; Guang-Uei Hung; Tao-Cheng Wu; Wen-Sheng Huang; Dianfu Li; Robert A deKemp; Bailing Hsu
Journal:  J Nucl Cardiol       Date:  2014-10-04       Impact factor: 5.952

2.  Influences of reconstruction and attenuation correction in brain SPECT images obtained by the hybrid SPECT/CT device: evaluation with a 3-dimensional brain phantom.

Authors:  Mana Akamatsu; Yasuo Yamashita; Go Akamatsu; Yuji Tsutsui; Nobuyoshi Ohya; Yasuhiko Nakamura; Masayuki Sasaki
Journal:  Asia Ocean J Nucl Med Biol       Date:  2014

3.  Importance of Defect Detectability in Positron Emission Tomography Imaging of Abdominal Lesions.

Authors:  Shozo Yamashita; Kunihiko Yokoyama; Masahisa Onoguchi; Haruki Yamamoto; Tetsu Nakaichi; Shiro Tsuji; Kenichi Nakajima
Journal:  Asia Ocean J Nucl Med Biol       Date:  2015

4.  Influence of reconstruction settings on the performance of adaptive thresholding algorithms for FDG-PET image segmentation in radiotherapy planning.

Authors:  Roberta Matheoud; Patrizia Della Monica; Gianfranco Loi; Luca Vigna; Marco Krengli; Eugenio Inglese; Marco Brambilla
Journal:  J Appl Clin Med Phys       Date:  2011-01-30       Impact factor: 2.102

5.  The effect of calibration factors and recovery coefficients on 177Lu SPECT activity quantification accuracy: a Monte Carlo study.

Authors:  Keamogetswe Ramonaheng; Johannes A van Staden; Hanlie du Raan
Journal:  EJNMMI Phys       Date:  2021-03-18

6.  Effects of the attenuation correction and reconstruction method parameters on conventional cardiac dynamic SPECT.

Authors:  Mohammadreza Mohseni; Reza Faghihi; Mahdi Haghighatafshar; Seyed Mohammad Entezarmahdi
Journal:  Medicine (Baltimore)       Date:  2018-09       Impact factor: 1.889

7.  Time-modified OSEM algorithm for more robust assessment of left ventricular dyssynchrony with phase analysis in ECG-gated myocardial perfusion SPECT.

Authors:  Matti J Kortelainen; Tuomas M Koivumäki; Marko J Vauhkonen; Mikko A Hakulinen
Journal:  EJNMMI Phys       Date:  2019-12-27

8.  Optimization of Ordered Subset Expectation Maximization Parameters for Image Reconstruction in Tc-99m Methoxyisobutylisonitrile Myocardial Perfusion SPECT and Comparison with Corresponding Filtered Back Projection-Reconstructed Images.

Authors:  Pankaj Dheer; Priyanka Gupta; Sameer Kamalakar Taywade; Averilicia Passah; Anil Kumar Pandey; Chetan Patel
Journal:  Indian J Nucl Med       Date:  2021-03-04
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.