Literature DB >> 15909193

A newly developed intra-operative gamma camera: performance characteristics in a laboratory phantom study.

Doron Kopelman1, Ira Blevis, Galina Iosilevsky, Alla Reznik, Alex Chaikov, Naor Weiner, Ora Israel, Moshe Hashmonai.   

Abstract

PURPOSE: Radioguided surgery depends on the intra-operative detection of radiolabelled tissues. This is currently accomplished with hand tools capable of providing a tone signal, depending on the proximity and direction of a radioactive source in relation to the probe. The advantages of visual images of radiolabelled tissues are well recognised, but satisfactory means of acquiring such images intra-operatively are not yet available. The goal of this study was to examine the performance of a newly developed intra-operative gamma camera, compact enough to be a hand tool and capable of yielding a visual image of the source field.
METHODS: The study was performed in the laboratory with a phantom consisting of a water bath and small hollow spheres (1-2 cm in internal diameter) filled with 99mTc (1-5 microCi/cc), placed in different configurations within the bath. For comparison, studies were also performed using a standard intra-operative gamma probe, and others using a standard single-head high-resolution gamma camera.
RESULTS: Compared with the gamma probe, the intra-operative camera was found to possess a superior ability to distinguish small, deep and weakly localised radioactivity sources from background. By acquiring images from different angles, it allowed a 3D understanding of multiple radioactive sources. It detected "cold" defects within a "hot" radiolabelled sphere. It discriminated a weak source located near a much "hotter" radioactivity source, similar to discrimination with the standard gamma camera, and discerned localised sources against a background of radioactivity.
CONCLUSION: It is anticipated that the high imaging potential of the camera tested in this study will offer clinical advantages.

Entities:  

Mesh:

Year:  2005        PMID: 15909193     DOI: 10.1007/s00259-005-1823-y

Source DB:  PubMed          Journal:  Eur J Nucl Med Mol Imaging        ISSN: 1619-7070            Impact factor:   9.236


  12 in total

1.  Radio-guided sentinel node detection for gastric cancer.

Authors:  Y Kitagawa; H Fujii; M Mukai; T Kubota; Y Otani; M Kitajima
Journal:  Br J Surg       Date:  2002-05       Impact factor: 6.939

2.  Evaluation of surgical gamma probes for radioguided sentinel node localisation.

Authors:  T Tiourina; B Arends; D Huysmans; H Rutten; B Lemaire; S Muller
Journal:  Eur J Nucl Med       Date:  1998-09

Review 3.  Sentinel lymph node detection and imaging.

Authors:  M R Keshtgar; P J Ell
Journal:  Eur J Nucl Med       Date:  1999-01

4.  A method to evaluate intra-operative gamma probes for sentinel lymph node localisation.

Authors:  A J Britten
Journal:  Eur J Nucl Med       Date:  1999-02

5.  Intraoperative lymph scintigraphy during radical surgery for cervical cancer.

Authors:  E Gitsch; K Philipp; N Pateisky
Journal:  J Nucl Med       Date:  1984-04       Impact factor: 10.057

Review 6.  Radioguided sentinel lymph node biopsy in breast cancer surgery.

Authors:  G Mariani; L Moresco; G Viale; G Villa; M Bagnasco; G Canavese; J Buscombe; H W Strauss; G Paganelli
Journal:  J Nucl Med       Date:  2001-08       Impact factor: 10.057

7.  Probe-guided surgery for colorectal cancer.

Authors:  P Lechner; P Lind; M Snyder; H Haushofer
Journal:  Recent Results Cancer Res       Date:  2000

8.  The sentinel node procedure with Patent Blue V dye in the surgical treatment of papillary thyroid carcinoma.

Authors:  M R Pelizzo; I M Boschin; A Toniato; P Bernante; A Piotto; A Rinaldo; A Ferlito
Journal:  Acta Otolaryngol       Date:  2001-04       Impact factor: 1.494

9.  The role of lymphoscintigraphy in rectal laparoscopic surgery: can the sentinel node concept be applied to rectal carcinoma?

Authors:  G Cutini; G C Gesuelli; M Sartelli; E Brianzoni; G Musolino; M Nestori; R Scibé; A Berbellini
Journal:  Surg Endosc       Date:  2001-12       Impact factor: 4.584

10.  Diagnostic use of the sentinel node in colon cancer.

Authors:  A E Merrie; A M van Rij; L V Phillips; J I Rossaak; K Yun; J L Mccall
Journal:  Dis Colon Rectum       Date:  2001-03       Impact factor: 4.585

View more
  5 in total

Review 1.  SPECT detectors: the Anger Camera and beyond.

Authors:  Todd E Peterson; Lars R Furenlid
Journal:  Phys Med Biol       Date:  2011-08-09       Impact factor: 3.609

2.  Quantitative investigation of a novel small field of view hybrid gamma camera (HGC) capability for sentinel lymph node detection.

Authors:  Mohammed S Alqahtani; John E Lees; Sarah L Bugby; Layal K Jambi; Alan C Perkins
Journal:  Br J Radiol       Date:  2016-08-18       Impact factor: 3.039

3.  Evaluation of two intraoperative gamma detectors for assessment of 177Lu activity concentration in vivo.

Authors:  Viktor Sandblom; Ingun Ståhl; Roger Olofsson Bagge; Eva Forssell-Aronsson
Journal:  EJNMMI Phys       Date:  2017-01-09

Review 4.  Above and Beyond Robotic Surgery and 3D Modelling in Paediatric Cancer Surgery.

Authors:  Laura Privitera; Irene Paraboschi; Kate Cross; Stefano Giuliani
Journal:  Front Pediatr       Date:  2021-12-20       Impact factor: 3.418

5.  Optimization of sentinel lymph node biopsy in breast cancer using an operative gamma camera.

Authors:  Carole Mathelin; Samuel Salvador; Sabrina Croce; Norosoa Andriamisandratsoa; Daniel Huss; Jean-Louis Guyonnet
Journal:  World J Surg Oncol       Date:  2007-11-17       Impact factor: 2.754