Literature DB >> 15908528

Detection and measurement of rotator cuff tears with sonography: analysis of diagnostic errors.

Sharlene A Teefey1, William D Middleton, William T Payne, Ken Yamaguchi.   

Abstract

OBJECTIVE: The purpose of this study was to analyze the causes of errors in the detection and measurement of rotator cuff tears in our patient population. SUBJECTS AND METHODS: Seventy-one consecutive patients with shoulder pain who were prospectively studied with sonography had subsequent arthroscopy that showed a full-thickness or partial-thickness tear or intact cuff. For sonography and arthroscopy, the length or degree of retraction and width of a tear, when present, was recorded. When there were discrepant findings, representative images were jointly evaluated by the radiologist and orthopedic surgeon to determine the cause of the error.
RESULTS: Fifteen detection errors were found, including five misses (three < 5-mm subscapularis and two small partial-thickness tears), four errors inherent with the test (distinguishing large bursal side or extensive partial-thickness from full-thickness tears and tendinopathy from partial-thickness tears), three errors of an unknown cause, two due to misinterpretation, and one error inherent with the patient. Seventeen measurement errors occurred with full-thickness tears, 15 of those in patients with large or massive tears. Bursal thickening (n = 4), non-visualization of the torn tendon end (n = 2), nonretracted tear (n = 2), and complex tear (n = 1) contributed to the errors. Eight measurement errors occurred with partial-thickness tears. Difficulty distinguishing tendinopathy from partial-thickness tears (n = 3) and complex tears (n = 3) accounted for six errors.
CONCLUSION: Although infrequent, detection errors were due to limitations inherent with the test or misses. Limitations inherent with the patient and misinterpretation of the findings were rare. Most measurement errors occurred in patients with large or massive cuff tears.

Entities:  

Mesh:

Year:  2005        PMID: 15908528     DOI: 10.2214/ajr.184.6.01841768

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  16 in total

Review 1.  Anterior-superior internal impingement of the shoulder: an evidence-based review.

Authors:  Raffaele Garofalo; Jon Karlsson; Ulf Nordenson; Eugenio Cesari; Marco Conti; Alessandro Castagna
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-08-11       Impact factor: 4.342

2.  Location and initiation of degenerative rotator cuff tears: an analysis of three hundred and sixty shoulders.

Authors:  H Mike Kim; Nirvikar Dahiya; Sharlene A Teefey; William D Middleton; Georgia Stobbs; Karen Steger-May; Ken Yamaguchi; Jay D Keener
Journal:  J Bone Joint Surg Am       Date:  2010-05       Impact factor: 5.284

3.  Surgeon-operated ultrasonography in a one-stop shoulder clinic.

Authors:  R Seagger; T Bunker; P Hamer
Journal:  Ann R Coll Surg Engl       Date:  2011-10       Impact factor: 1.891

4.  Rotator cuff tears: assessment with MR arthrography in 275 patients with arthroscopic correlation.

Authors:  S Waldt; M Bruegel; D Mueller; K Holzapfel; A B Imhoff; E J Rummeny; K Woertler
Journal:  Eur Radiol       Date:  2006-09-13       Impact factor: 5.315

Review 5.  Comprehensive Shoulder US Examination: A Standardized Approach with Multimodality Correlation for Common Shoulder Disease.

Authors:  Matthew H Lee; Scott E Sheehan; John F Orwin; Kenneth S Lee
Journal:  Radiographics       Date:  2016-10       Impact factor: 5.333

6.  US appearance of partial-thickness supraspinatus tendon tears: Application of the string theory. Pictorial essay.

Authors:  H Guerini; M Fermand; D Godefroy; A Feydy; A Chevrot; G Morvan; N Gault; J L Drapé
Journal:  J Ultrasound       Date:  2012-01-15

7.  Relationship of tear size and location to fatty degeneration of the rotator cuff.

Authors:  H Mike Kim; Nirvikar Dahiya; Sharlene A Teefey; Jay D Keener; Leesa M Galatz; Ken Yamaguchi
Journal:  J Bone Joint Surg Am       Date:  2010-04       Impact factor: 5.284

8.  Diagnostic accuracy of ultrasound in subscapularis tendon abnormalities and the importance of operator experience.

Authors:  Uğur Toprak; Sefa Türkoğlu; Çiğdem Aydoğan; Emrah Kovalak; Suzan Saylısoy; Duygu Sıddıkoğlu; Akkan Mahmud
Journal:  Acta Orthop Traumatol Turc       Date:  2020-07       Impact factor: 1.511

9.  Classification of rotator cuff tendinopathy using high definition ultrasound.

Authors:  Hannah Hinsley; Alex Nicholls; Michael Daines; Gemma Wallace; Nigel Arden; Andrew Carr
Journal:  Muscles Ligaments Tendons J       Date:  2014-11-17

10.  Sonography of the painful shoulder: role of the operator's experience.

Authors:  Thomas Le Corroller; Michel Cohen; Richard Aswad; Vanessa Pauly; Pierre Champsaur
Journal:  Skeletal Radiol       Date:  2008-07-24       Impact factor: 2.199

View more